Z-2285-2012Class III

PHADIA Varelisa Cardiolipin IgM Cardiolipin autoantibody immunological test system (IgM class antibodies)

Phadia AB / initiated 2012-08-02 / Terminated / root cause: labeling and UDI

Phadia US Inc began a recall of PHADIA Varelisa Cardiolipin IgM Cardiolipin autoantibody immunological test system (IgM class antibodies) on . FDA classified it as Class III and gives the reason as "Customers were not following the labeled assay procedure. This was due to a letter issued to them in 2004 from Phadia GmbH, which indicated that customers could omit the pre-wash step. The omission of the pre-wash step...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2285-2012
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Phadia AB
Product
PHADIA Varelisa Cardiolipin IgM Cardiolipin autoantibody immunological test system (IgM class antibodies)
Product code
MID System, Test, Anticardiolipin Immunological
Reason for recall
Customers were not following the labeled assay procedure. This was due to a letter issued to them in 2004 from Phadia GmbH, which indicated that customers could omit the pre-wash step. The omission of the pre-wash step, when combined with a change in material source in kit lot #0067, caused elevated results.
Root cause tag
labeling and UDI
FDA root cause
Labeling Change Control
Action
On 8/2/2012 an URGENT PRODUCT CORRECTION INFORMATION FOR PHADIA VARELISA CARDIOLIPIN IGM, Article Number 15696 letter was sent to all consignees the received the Direction for Use modification letter in 2004.
Quantity in commerce
837 kits (96 assays per kit), total of 80,352 individual assays
Distribution
USA Nationwide.
Product codes and lots
Article Number 15696 Lot number #0067

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2285-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2285-2012Class IIITerminatedVoluntary: Firm initiated