Z-2209-2012Class II

COBAS AmpliPrep Sample Tube Input Barcode Clips, catalog number 03137040001. Input S-tube barcode clips hold...

Roche Molecular Systems Inc / initiated 2012-02-01 / Terminated / root cause: manufacturing process

Roche Molecular Systems, Inc. began a recall of COBAS AmpliPrep Sample Tube Input Barcode Clips, catalog number 03137040001. Input S-tube barcode clips hold sample input tubes in the COBAS AmpliPrep... on . FDA classified it as Class II and gives the reason as "Certain lots of COBAS AmpliPrep (CAP) S-tube Input (Box of 12 bags 12x24 S-tube and 12x24 S-clip) can cause the sample input tubes to stick in the barcode clips. As a consequence, the CAP instrument will lift both the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2209-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Roche Molecular Systems Inc
Product
COBAS AmpliPrep Sample Tube Input Barcode Clips, catalog number 03137040001. Input S-tube barcode clips hold sample input tubes in the COBAS AmpliPrep Instrument (CAP) sample racks and have a barcode to identify the sample.
Product code
JJF Analyzer, Chemistry, Micro, For Clinical Use
Reason for recall
Certain lots of COBAS AmpliPrep (CAP) S-tube Input (Box of 12 bags 12x24 S-tube and 12x24 S-clip) can cause the sample input tubes to stick in the barcode clips. As a consequence, the CAP instrument will lift both the sample input tube together with the barcode clip from the sample rack, resulting in an aborted run. No erroneous results would be produced if the issue occurs, but samples in proce
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On 1/17/2012 an "Analyzer Bulletin" notification was sent to Roche affiliates and on 2/1/2012 an "Analyzer Bulletin" notification was sent to all affected US Consignees. The notification described the product issue and provided recommended actions. Customers were referred to Roche Technical Support at 800-526-1247 for questions.
Quantity in commerce
2,132 boxes
Distribution
Nationwide Distribution (USA).
Product codes and lots
Lot Numbers: 11049503, 11053503, 11054503, 11525503 and 11526503.