Z-2215-2012Class II

00-4309-029-01 - Trabecular Metal Reverse Shoulder Liner 36 mm 65 Neck Angle Per surgical technique...

Zimmer GmbH / initiated 2012-07-18 / Terminated / root cause: design

Zimmer, Inc. began a recall of 00-4309-029-01 - Trabecular Metal Reverse Shoulder Liner 36 mm 65 Neck Angle Per surgical technique 97-4309-003-00 Revision 4: "Complete Poly Liner insertion... on . FDA classified it as Class II and gives the reason as "Lot specific Trabecular Metal (TM) Reverse Shoulder Liner Impactors are being recalled due to reports of alignment peg fractures. All lots of TM Reverse Shoulder Liner Impactors manufactured from 455 stainless steel are...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2215-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
00-4309-029-01 - Trabecular Metal Reverse Shoulder Liner 36 mm 65 Neck Angle Per surgical technique 97-4309-003-00 Revision 4: "Complete Poly Liner insertion by snapping the appropriate 60 (Standard) or 65 (Retentive) Poly Liner Impactor to the Poly Liner Impactor Handle. Then, place the Poly Liner Impactor on the Poly Liner articular surface and forcefully strike the Poly Liner Handle with a mallet."
Product code
KWA Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component)
Reason for recall
Lot specific Trabecular Metal (TM) Reverse Shoulder Liner Impactors are being recalled due to reports of alignment peg fractures. All lots of TM Reverse Shoulder Liner Impactors manufactured from 455 stainless steel are being retrieved, as all reported fractured instruments were manufactured with this material. The TM Reverse Shoulder Liner Impactor units that were manufactured with 13-8 stain
Root cause tag
design
FDA root cause
Device Design
Action
Zimmmer sent an Urgent Medical Device Recall letter dated July 18, 2012, to all affected customers. The letter identified the product, the problem, and thea action to be taken by the customer. Customers were instructed to inspect all TM Reverse instrument sets for affected lot(s)(instruments. If affected product was found customers were instructed to immediately place an order to backfill the identified product with Customer Service. If an instrument was was further distributed customers were to provide a copy of the Notice to ensure their customers were aware of the removal. Customers were also instructed to provide a list of all hospitals that currently use the affected product. For questions customers were instructed to call 800-348-2759. For questions regarding this recall call 574-371-1689.
Quantity in commerce
212 (179 USA)
Distribution
Worldwide Distribution - USA (nationwide) and internationally to Germany, Belgium, Denmark, Finland, France, Italy, Netherlands, Spain, Sweden, Switzerland, and the United Kingdom.
Product codes and lots
Part 00-4309-029-01: Lot 60535162, 60545338, 60576488,  60605855, 60637022, 60689253, 60772974, 60875277, 60918060, 60989817, 61011689, 61024161, and 61051024.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2215-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2212-2012Class IITerminatedVoluntary: Firm initiated
Z-2213-2012Class IITerminatedVoluntary: Firm initiated
Z-2214-2012Class IITerminatedVoluntary: Firm initiated
Z-2215-2012Class IITerminatedVoluntary: Firm initiated