Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2144-2012Class II

Smith & Nephew Dyonics Electroblade Resector, 4.5mm Full Radius Blade, BONECUTTER Part Number: 72202213

Smith & Nephew, Inc. / initiated 2012-07-14 / Terminated / root cause: labeling and UDI

Smith & Nephew, Inc. Endoscopy Division began a recall of Smith & Nephew Dyonics Electroblade Resector, 4.5mm Full Radius Blade, BONECUTTER Part Number: 72202213 Product Usage: The Dyonics Electroblade Resector is... on . FDA classified it as Class II and gives the reason as "Devices labeled with an expiry period of 48 months instead of 36 months; an additional 12 months for which product and package reliability has not been validated". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2144-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
Smith & Nephew Dyonics Electroblade Resector, 4.5mm Full Radius Blade, BONECUTTER Part Number: 72202213 Product Usage: The Dyonics Electroblade Resector is indicated for use in arthroscopic surgical procedures of large and small articular cavities. The Dyonics Electroblade Resector is effective in soft and osseous tissue resection and in hemostasis of bleeding vessels. It is intended for arthroscopic procedures using saline solution, Ringers lactate or other conductive solutions as an irrigant under direct or video assisted fiberoptic visualization.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Devices labeled with an expiry period of 48 months instead of 36 months; an additional 12 months for which product and package reliability has not been validated.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Smith & Nephew sent an Urgent - Product Recall 1st Notification Letter dated July 2012 via Federal Express to all affected customers. The letter identified the affected product, potential risk and actions to be taken. Customers were instructed to examine inventory, quarantine immediately, complete the form attached indicating the lot numbers and quantities. The letter also instructed customers to contact Smith & Nephew Returns Group at 800-343-5717 (option 3) for a Return Authorization (RA) or send an email to [email protected]. They will provide customers with instructions for returning affected products and receiving a credit. Each OUS affiliate shall be notified via (email or equivalent) of the recall by Smith & Nephew Inc.; the affiliate must then notify each facility affected by the recall. For question concerning this recall call 405-917-8585 or 978-749-1630.
Quantity in commerce
340 units
Distribution
Worldwide Distribution - US Nationwide and the countries of : Australia, Canada, Korea, Turkey, UAE, and Japan.
Product codes and lots
Lot Numbers: 950135R, 950136R, 950137R, 950138R

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2144-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2144-2012Class IITerminatedVoluntary: Firm initiated
Z-2145-2012Class IITerminatedVoluntary: Firm initiated