Z-2182-2012Class II

Dimension(R) Reagent Management System (RMS) - Power Cord. The Siemens Healthcare Diagnostics Dimension(R)...

Siemens Healthcare Diagnostics Inc. / initiated 2012-07-17 / Terminated / root cause: design

Siemens Healthcare Diagnostics, Inc. began a recall of Dimension(R) Reagent Management System (RMS) - Power Cord. The Siemens Healthcare Diagnostics Dimension(R) clinical chemistry analyzers are floor model, fully... on . FDA classified it as Class II and gives the reason as "Siemens Healthcare Diagnostics has confirmed that partial insertion of the power cord on the Dimension Reagent Management System (RMS) power input module can cause the plug to overheat and potentially create a fire...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2182-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Dimension(R) Reagent Management System (RMS) - Power Cord. The Siemens Healthcare Diagnostics Dimension(R) clinical chemistry analyzers are floor model, fully automated, microprocessor-controlled, integrated instrument systems that use prepackaged Flex(R) reagent test cartridges to measure a variety of analytes in human body fluids.
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Siemens Healthcare Diagnostics has confirmed that partial insertion of the power cord on the Dimension Reagent Management System (RMS) power input module can cause the plug to overheat and potentially create a fire hazard. RMS without a power cord retention clip may be affected.
Root cause tag
design
FDA root cause
Device Design
Action
Siemens sent an "URGENT FIELD SAFETY NOTICE" dated June 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Customers were asked to complete and return an attached Field Correction Effectiveness Check Form and fax it to (302) 631-8467. Contact Siemens Technical Solutions Center at 800-441-9250 for questions regarding this notice.
Quantity in commerce
2,554
Distribution
Distributed in the US including Washington, DC, Puerto Rico and the states of AL, AK, AZ, AR, CA, CO, CT, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, VA, WA, WI, WV, and WY and the Virgin and Pacific Islands.
Product codes and lots
Serial numbers 99060001 - 99062554

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2182-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2182-2012Class IITerminatedVoluntary: Firm initiated