Z-2207-2012Class II

Philips Essenta DR, X-Ray System, Diagnostic, general-purpose Model: 712070. The Essenta DR is a...

Philips North America LLC / initiated 2012-07-25 / Terminated / root cause: manufacturing process

Philips Healthcare Inc. began a recall of Philips Essenta DR, X-Ray System, Diagnostic, general-purpose Model: 712070. The Essenta DR is a multifunctional stationary X-ray Intended Use From Labeling)... on . FDA classified it as Class II and gives the reason as "Essenta DR C-Arm may fall". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2207-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Philips Essenta DR, X-Ray System, Diagnostic, general-purpose Model: 712070. The Essenta DR is a multifunctional stationary X-ray Intended Use From Labeling): system, in which the swivel arm rotation, the vertical movement of the swivel arm and the SID adjustment are motorized. It is designed for radiographic examination of the recumbent, standing or seated patient and also for lateral exposures. It enables the operator to quickly and safely move the unit to all the pre-programmed basic positions at the press of a button, using the autopositioning feature. Fine positioning of the tube/collimator and detector on the patient is easy to carry out using the command arm.
Product code
KPR System, X-Ray, Stationary
Reason for recall
Essenta DR C-Arm may fall
Root cause tag
manufacturing process
FDA root cause
Equipment maintenance
Action
Philips Medical sent an "URGENT FIELD SAFETY NOTICE" dated July 09, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. Customers are asked to inspect the C-Arm unit and discontinue use if not parallel to the column, do not use the system and call for a Service Technician. A Field Change Order (FCO) is being released that requires an inspection of C-ARM connection and, if necessary, repair of systems. The bolts will be disconnected and thread inserts will be checked. If not the thread inserts are no longer secured, additional holes will be drilled to use in place of the original ones. For further information or support concerning this issue, contact your local Philips representative at 1-800-722-9377.
Quantity in commerce
7 units
Distribution
Worldwide Distribution-USA including the states of CA, NC, and TN and the countries of China, Germany, and Hungary.
Product codes and lots
Serial Numbers:  SN09000006/ SN09000007/ SN10000033/ SN10000014/ SN100000654/ SN08000092/ SN08000050

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2207-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2207-2012Class IITerminatedVoluntary: Firm initiated