Z-2264-2012Class II

syngo Imaging XS. Radiological image processing system

Siemens Medical Solutions USA, Inc. / initiated 2012-07-18 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of syngo Imaging XS. Radiological image processing system on . FDA classified it as Class II and gives the reason as "There is a potential malfunction when using syngo Imaging XS, version VA70A or higher. In some situations it can happen that a miscalculation is done for the grey scale values in the functions "Regions of Interest"...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2264-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
syngo Imaging XS. Radiological image processing system.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
There is a potential malfunction when using syngo Imaging XS, version VA70A or higher. In some situations it can happen that a miscalculation is done for the grey scale values in the functions "Regions of Interest", "Pixel Lens", "Edge Enhancement" and "Histograms".
Root cause tag
software
FDA root cause
Software design
Action
Siemens sent a "CUSTOMER SAFETY ADVISORY NOTICE" dated July 12, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. For questions concerning this notice contact your local Siemens Service Rep..
Quantity in commerce
52
Distribution
Nationwide Distribution-CA, CO, CT, FL, GA, IL, IN, IA, KS, MD, MN, MO, NE, NV, NJ, NY, NC, OH, PA, TN, TX, and WI.
Product codes and lots
Model number 10496279

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2264-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2264-2012Class IITerminatedVoluntary: Firm initiated