Z-2275-2012Class II
syngo.plaza, Model number 10592457
Siemens Medical Solutions USA, Inc. / initiated 2012-07-18 / Terminated / root cause: software
Siemens Medical Solutions USA, Inc began a recall of syngo.plaza, Model number 10592457. Intended use: radiological image processing system on . FDA classified it as Class II and gives the reason as "There is a potential malfunction in which a miscalculation occurs for images with pixel depth values greater that 12 bit in certain functions". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2275-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- syngo.plaza, Model number 10592457. Intended use: radiological image processing system.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- There is a potential malfunction in which a miscalculation occurs for images with pixel depth values greater that 12 bit in certain functions.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm initiated their recall/correction of this device on July 18, 2012 by sending a "Customer Safety Advisory Notice" letter, dated July 12, 2012, to their consignees. The letter identified the affected product along with the malfunction. It also discussed the steps the user can take to avoid the issue. Users were asked to not use the mentioned functions for images from those modalities until the issue was resolved and the system was updated with the software version containing the solution. Users were to contact their local Siemens Service in order to plan for the updated software version that will resolve the issue. Additionally, users should place the safety advisory notice in the system's instructions for use, and personnel should be instructed accordingly and maintain awareness. If the device was sold and is no longer in the users' possession, the notice should be forwarded to the new owner and the firm should be notified of the new owner.
- Quantity in commerce
- 14
- Distribution
- Nationwide Distribution -- CA, FL, GA, MA, MN, MO, NJ, NY, OH, WA, and WI.
- Product codes and lots
- Serial Numbers: 100263, 100401, 100177, 100138, 100174, 100219, 100406, 100249, 200040, 100201, 100198, 100181, 100301, & 100146.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2275-2012 | Class II | Terminated | Voluntary: Firm initiated |