Z-2275-2012Class II

syngo.plaza, Model number 10592457

Siemens Medical Solutions USA, Inc. / initiated 2012-07-18 / Terminated / root cause: software

Siemens Medical Solutions USA, Inc began a recall of syngo.plaza, Model number 10592457. Intended use: radiological image processing system on . FDA classified it as Class II and gives the reason as "There is a potential malfunction in which a miscalculation occurs for images with pixel depth values greater that 12 bit in certain functions". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2275-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
syngo.plaza, Model number 10592457. Intended use: radiological image processing system.
Product code
LLZ System, Image Processing, Radiological
Reason for recall
There is a potential malfunction in which a miscalculation occurs for images with pixel depth values greater that 12 bit in certain functions.
Root cause tag
software
FDA root cause
Software design
Action
The firm initiated their recall/correction of this device on July 18, 2012 by sending a "Customer Safety Advisory Notice" letter, dated July 12, 2012, to their consignees. The letter identified the affected product along with the malfunction. It also discussed the steps the user can take to avoid the issue. Users were asked to not use the mentioned functions for images from those modalities until the issue was resolved and the system was updated with the software version containing the solution. Users were to contact their local Siemens Service in order to plan for the updated software version that will resolve the issue. Additionally, users should place the safety advisory notice in the system's instructions for use, and personnel should be instructed accordingly and maintain awareness. If the device was sold and is no longer in the users' possession, the notice should be forwarded to the new owner and the firm should be notified of the new owner.
Quantity in commerce
14
Distribution
Nationwide Distribution -- CA, FL, GA, MA, MN, MO, NJ, NY, OH, WA, and WI.
Product codes and lots
Serial Numbers: 100263, 100401, 100177, 100138, 100174, 100219, 100406, 100249, 200040, 100201, 100198, 100181, 100301, & 100146.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2275-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2275-2012Class IITerminatedVoluntary: Firm initiated