Z-2240-2012Class II

892.1650 Image-intensified fluoroscopic x-ray system; the device is designed to provide fluoroscopic and spot...

GE Healthcare / initiated 2009-11-06 / Terminated / root cause: design

GE OEC Medical Systems, Inc began a recall of 892.1650 Image-intensified fluoroscopic x-ray system; the device is designed to provide fluoroscopic and spot film imaging of the patient during diagnostic... on . FDA classified it as Class II and gives the reason as "GE OEC is recalling certain 9800 Image Intensified Flouroscopic x-ray systems due to customer complaint analysis showing that a replacement Hitachi 160 GB Hard Drive, when installed on certain models is susceptible of...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2240-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
892.1650 Image-intensified fluoroscopic x-ray system; the device is designed to provide fluoroscopic and spot film imaging of the patient during diagnostic surgical and interventional procedures. The systems include features specifically designed for use in diagnostic and interventional cardiac imaging procedures, and are also intended for cholangiography, endoscopic, urologic, orthopedic, neurologic, vascular, critical care and emergency room procedures. The systems may be used for other imaging applications at the physician's discretion.
Product code
JAA System, X-Ray, Fluoroscopic, Image-Intensified
Reason for recall
GE OEC is recalling certain 9800 Image Intensified Flouroscopic x-ray systems due to customer complaint analysis showing that a replacement Hitachi 160 GB Hard Drive, when installed on certain models is susceptible of causing unanticipated system shut downs, no boots, data loss or data mix, or unexpected hard drive corruption.
Root cause tag
design
FDA root cause
Device Design
Action
GE OEC sent a Urgent Recall Notice dated November 6, 2009, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The notification advised them that GE Healthcare had identified a potential safety issue associated with a "Significant increase on Data Loss, Data Mix, and System file corruption of the Hard Drives" on the 9800 Systems. The field correction consisted of installing a "Smart Power Switch" between the power control PCB and the system power switch on 9800 Systems with a Celeron single board computer and software version 29. This will remedy the potential for data loss, data mix, and system file corruption of the hard drives. For further questions please call 800-874-7378 Option 8.
Quantity in commerce
3,595 units
Distribution
Worldwide Distribution -- USA (nationwide)
Product codes and lots
Model Number HDS721616PLAT80; GE Healthcare part number 5304649.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2240-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2240-2012Class IITerminatedVoluntary: Firm initiated