syngo(R) Lab Data Manager and syngo(R) Lab Process Manager, all versions. This device is classified as a data...
Siemens Healthcare Diagnostics, Inc. began a recall of syngo(R) Lab Data Manager and syngo(R) Lab Process Manager, all versions. This device is classified as a data calculator and processing module for use with... on . FDA classified it as Class II and gives the reason as "When using the syngo Quality Control package the system may not perform as intended for Multi-Rule QC violation [e.g., 2(2S), 4(1s) and 10x], unless configured using the instructions on page 2 of the recall...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2305-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- syngo(R) Lab Data Manager and syngo(R) Lab Process Manager, all versions. This device is classified as a data calculator and processing module for use with In-vitro diagnostic devices. It also provides work flow management as well as remote instrument control to multiple interfaced instruments.
- Product code
- JQP Calculator/Data Processing Module, For Clinical Use
- Reason for recall
- When using the syngo Quality Control package the system may not perform as intended for Multi-Rule QC violation [e.g., 2(2S), 4(1s) and 10x], unless configured using the instructions on page 2 of the recall notification. QC status may not be flagged appropriately. If auto-verification rules are defined to hold patient results when QC status is out, the patient results will not be held as expecte
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Siemens sent a Urgent Field Safety Notice letter dated May, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter instructed the customers to re-set the Multi-Rule option and instructions are provided on how to do so. This recall notification also included a Field Safety Effectiveness check form to please fax to the Siemens Technical Solutions Center at 302-631-8467 BY June 15th.
- Quantity in commerce
- 44
- Distribution
- USA ( nationwide ) including the states of AZ, CA, FL, IL, KY, MD, MA, MN, MS, MO, NE, NH, NJ, NY, NC, OH, PA, TX, VA and WV.
- Product codes and lots
- Version VA11B and all previously released versions.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2305-2012 | Class II | Terminated | Voluntary: Firm initiated |