Siemens Stratus(R) CS Acute Care(TM) D-dimer Calibrator (DDM CalPak) lots 502058002 and 501318002 The...
Siemens Healthcare Diagnostics, Inc. began a recall of Siemens Stratus(R) CS Acute Care(TM) D-dimer Calibrator (DDM CalPak) lots 502058002 and 501318002 The Stratus(R) CS Acute Care(TM) D-dimer Calibrator (DDMR... on . FDA classified it as Class III and gives the reason as "Siemens has confirmed that the Stratus(R) CS Acute Care(TM) D-Dimer CalPak calibrator lots 502058002 and 501318002 may exhibit SLOPE or CALCV error messages (instrument errors), which indicate calibration failures". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2283-2012
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Siemens Stratus(R) CS Acute Care(TM) D-dimer Calibrator (DDM CalPak) lots 502058002 and 501318002 The Stratus(R) CS Acute Care(TM) D-dimer Calibrator (DDMR CalPak), Catalog No. CDDMR-C is an in-vitro diagnostic product intended to be used for the calibration of the Stratus(R) CS Acute Care(TM) D-dimer )DDMR) method
- Product code
- DAP Fibrinogen And Fibrin Split Products, Antigen, Antiserum, Control
- Reason for recall
- Siemens has confirmed that the Stratus(R) CS Acute Care(TM) D-Dimer CalPak calibrator lots 502058002 and 501318002 may exhibit SLOPE or CALCV error messages (instrument errors), which indicate calibration failures.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- SIEMENS sent an Urgent Field Safety Notice dated June 7, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The letter included a response form to be faxed to the recalling firm at 302-631-8467 by the consignee. If customers experienced calibration failures, they were instructed to contact the Siemens Technical Solutions Center at 800-405-6473 for assistance. Customers were also asked to forward the notification to anyone to whom they may have distributed the affected product. For questions regarding this recall call 302-631-6311.
- Quantity in commerce
- 786 cartons
- Distribution
- Nationwide Distribution - USA including AK, AL, AR, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT, WA, WI, WV, and WY.
- Product codes and lots
- DDMR CalPak lots 502058002, exp 2012/11/23 and 501318002, exp 2012/08/10
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2283-2012 | Class III | Terminated | Voluntary: Firm initiated |