System 8000 Base
Terumo Cardiovascular Systems Corporation began a recall of System 8000 Base Product Usage: The Sarns Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for... on . FDA classified it as Class III and gives the reason as "During a retrospective review of complaint events, TCVS identified on 06/15/12 that three (3) Sarns" System 8000 bases were manufactured and distributed from 11/13/06-11/29/06 with an incorrect incorrect amperage rating...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2218-2012
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Product
- System 8000 Base Product Usage: The Sarns Modular Perfusion System 8000 is indicated for use for up to 6 hours on the extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures, when used by a qualified medical professional who is experienced in the operation of this or similar equipment.
- Product code
- DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
- Reason for recall
- During a retrospective review of complaint events, TCVS identified on 06/15/12 that three (3) Sarns" System 8000 bases were manufactured and distributed from 11/13/06-11/29/06 with an incorrect incorrect amperage rating identified on the name plate label. The mislabeling was identified in January 2007 a set up process of the automark label printer. All units were corrected in the field in Marc
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- Terumo field correction/service records were completed in 2007 and all affected units were corrected in the field in March 2007 by replacing the incorrect label at the customer location. For questions about this report call 734-741-6056.
- Quantity in commerce
- 3
- Distribution
- US Nationwide Distribution - including the state of New York
- Product codes and lots
- catalog number: 16401 and serial number: 2001, 2002, and 2003.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2218-2012 | Class III | Terminated | Voluntary: Firm initiated |