Z-2164-2012Class II

ABL90 FLEX software version 2.7 MR3. The ABL90 FLEX analyzer is intended for in vitro testing of samples of...

Radiometer America, Inc / initiated 2012-07-05 / Terminated / root cause: software

Radiometer America Inc began a recall of ABL90 FLEX software version 2.7 MR3. The ABL90 FLEX analyzer is intended for in vitro testing of samples of heparinized whole blood on . FDA classified it as Class II and gives the reason as "A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2164-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Radiometer America, Inc
Product
ABL90 FLEX software version 2.7 MR3. The ABL90 FLEX analyzer is intended for in vitro testing of samples of heparinized whole blood.
Product code
CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Reason for recall
A customer has reported an error on an ABL8xx configured to a Hospital Information System using the data transfer protocol HL7 version 2.5. If the instrument transmits results to a HIS system and the connection is broken the results are queued for later transmission. If the ABL and HIS system are configured with HL7 2.5, any results in the queue will be mixed up when the instrument transmits t
Root cause tag
software
FDA root cause
Software design
Action
The firm, Radiometer, sent two "Priority Level: Urgent" letters one dated July 9, 2012 and the other dated July 24, 2012 to its affected customers via FedEx with tracking. The letter identifies the product, problem, and actions to be taken by the customers. The customers were instructed to complete and return the attached Recall Response Fax Form via fax to 800-736-0601. The letter informed the customers that a local service representative will be contacting them shortly to install a software update that will prevent this from occuring (in the event that you utilize HL7 version 2.5 or intend to at some future date.) If you have any questions regarding this letter or the Recall, please contact Radiometer America Technical Support at 1-800-736-0600 opt. 4.
Quantity in commerce
1427 ABL 90 Instruments
Distribution
Worldwide distribution: USA (nationwide) including the states of AK, AL, AR, AZ, BM, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY, and the countries of Canada, Algeria, Argentina, Australia, Austria, Bahrain, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Columbia, Croatia, Cuba, Cyprus, Czech republic, Denmark, Ecuador, Egypt, Eritrea, Estonia, Finland, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Malaysia, Malta, Mauritius, Mexico, Moldova, Nepal, Netherlands, New Zealand, Norway, Oman, Pakistan, Peru, Philippines, Poland, Puerto Rico, Qatar, Republic of Korea, Romania, Russia, Saudi Arabia, Switzerland, Singapore, Slovenia, South Africa, Spain, Sweden Syria, Taiwan, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam, Yugoslavia, and Zimbabwe.
Product codes and lots
Part Number: 393-090, all production runs, 393-090R0026N0010 to 393-090R0179N010

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2164-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2162-2012Class IITerminatedVoluntary: Firm initiated
Z-2163-2012Class IITerminatedVoluntary: Firm initiated
Z-2164-2012Class IITerminatedVoluntary: Firm initiated