GEHC OEC Injection Cable for MedRad Provis V Contrast Injector. In certain fluoroscope X-ray modes, the cable...
GE OEC Medical Systems, Inc began a recall of GEHC OEC Injection Cable for MedRad Provis V Contrast Injector. In certain fluoroscope X-ray modes, the cable provides a signal to the auto-injector to trigger... on . FDA classified it as Class II and gives the reason as "GE initiated a correction to the instructions/use due to the potential for erroneous auto injection of contrast dye". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2262-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- GEHC OEC Injection Cable for MedRad Provis V Contrast Injector. In certain fluoroscope X-ray modes, the cable provides a signal to the auto-injector to trigger a contrast injection. Part numbers include 00-902260-01 and 00-883654-01.
- Product code
- OXO Image-Intensified Fluoroscopic X-Ray System, Mobile
- Reason for recall
- GE initiated a correction to the instructions/use due to the potential for erroneous auto injection of contrast dye.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- An Urgent Safety Advisory Notice was sent to consignees on October 10, 2008. The letter identified the affected product and notified customers of the situation, along with instructing them to discontinue use until proper validation and instruction update for cable use was completed. GEHC OEC was to have contacted customers when the solution became available and resolved it at no cost to the customer.
- Quantity in commerce
- 149 units
- Distribution
- Nationwide distribution.
- Product codes and lots
- Serial Numbers: 82-1138, 82-1586, 82-2089, 82-2117, 82-2182, 82-2214, 82-2226, 82-2262, 82-2264, 82-2308, 82-2312, 82-2314, 82-2321, 82-2328, 82-2380, 82-2391, 82-2438, 82-2483, 82-2512, 82-2516, 82-2535, 82-2722, 82-2796, 82-2821, 82-2846, 82-2848, 82-3052, 82-3055, 82-3129, 82-3144, 82-3170, 82-3171, 82-3201, 82-3256, 82-3267, 82-3314, 82-3350, 82-3361, 82-3377, 82-3467, 82-3468, 82-3659, 82-3709, 82-3722, 82-3859, 82-7016-MH, 82-7022-MH, 82-7037-MH, 82-7055-MH, 82-7062-MH, 82-7080-MH, 82-7086-MH, 82-7105-MH, 82-7114-MH, 82-7124-MH, 82-7131-MH, 82-7133-MH, 82-7137-MH, 82-7141-MH, 82-7154-CMH, 82-7155-MH, 82-7156-MH, 82-7161-MH, 82-7175-MH, 82-7189-MH, 82-7190-MH, 89-1441, 8S-1053, 8S-1436, 8S-2059-C, 8S-7021-CMH, 8S-7023-CMH, 8S-7040-CMH, 8S-7087-CMH, 8S-7164-CMH, DEMO/8S-1355-N, DEMO/E3-7002-MH, E2-0049, E2-0088, E2-0111, E2-0262, E2-7001-MH, E2-7007-MH, E2-7010-MH, E2-7014-MH, E2-7015-MH, E2-7031-MH, E2-7032-MH, E2-7035-MH, E2-7043-MH, E2-7059-MH, E2-7061-MH, E2-7067-MH, E2-7075-MH, E2-7076-MH, E2-7015-MH, & rENTAL/82-7004-MH.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2262-2012 | Class II | Terminated | Voluntary: Firm initiated |