Z-2179-2012Class II
Medstream 81" (206 cm) 20 Drop Standard bore Universal Administration Set with Y-Site, Sterile Manufactured...
Medical Specialties Distributors, Inc. / initiated 2012-08-03 / Terminated / root cause: manufacturing process
Medical Specialties Distributors, Inc. began a recall of Medstream 81" (206 cm) 20 Drop Standard bore Universal Administration Set with Y-Site, Sterile Manufactured for MSD (Medical Specialties Distributors LLC... on . FDA classified it as Class II and gives the reason as "Devices leaking at the Y-site". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2179-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medical Specialties Distributors, Inc.
- Product
- Medstream 81" (206 cm) 20 Drop Standard bore Universal Administration Set with Y-Site, Sterile Manufactured for MSD (Medical Specialties Distributors LLC, Stoughton, MA. Product of China Product Number: MS721 Disposable Infusion set
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- Devices leaking at the Y-site
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Medical Specialties Distributors LLC sent an Urgent Medical Recall letter dated August 2, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to inspect their inventory and remove all affected product from use immediately. Further distribution of the affected product should be discontinued immediately. Customers were instructed to contact their customers and inform them of the recall; complete and return the acknowledgement form and fax to 781-344-8320 to arrange for a return and to contact the Customer Service Department at (800) 967-6400 with questions. For questions regarding this recall call 781-344-6000.
- Quantity in commerce
- 472 cases (50/case)
- Distribution
- Nationwide Distribution
- Product codes and lots
- LOT NUMBER: 20111115
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2179-2012 | Class II | Terminated | Voluntary: Firm initiated |