Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2327-2012Class II

Bard Foley procedural trays packaged in breather bags. The product is packaged ten units in each shipper box...

C.R. Bard Inc / initiated 2012-06-29 / Terminated / root cause: manufacturing process

C.R. Bard, Inc., Urological Division began a recall of Bard Foley procedural trays packaged in breather bags. The product is packaged ten units in each shipper box, and are sold sterile for single use. Used to... on . FDA classified it as Class II and gives the reason as "Incomplete seal on sterile packages of product". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2327-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
C.R. Bard Inc
Product
Bard Foley procedural trays packaged in breather bags. The product is packaged ten units in each shipper box, and are sold sterile for single use. Used to catheterize patients. The finished trays are sold in Tyvek breather bags.
Product code
EZC Catheter, Coude
Reason for recall
Incomplete seal on sterile packages of product.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
C. R. Bard Inc, sent a Urgent Product Recall Notification letter dated June 27. 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. This notification is being sent because there is a potential for a limited number of units within the subject lots (noted above) that may contain an open seal of the sterile packages. The issue is visually obvious even before the package is opened. No patient injury or adverse events have been reported. Do not use or further distribute any of the affected products. Once the product affected by tlle recall has been removed from your inventory 1.Fill out the Recall & Effectiveness Check Form. Be sure to state the quantities affected of the recalled product that you have in stock. It is extremely important that we receive this information. 2. Fax the Recall & Effectiveness Check Form to BMD at 1-770-784-6469. The Recall Coordinator will issue you with a Return Authorization (RCL) to facilitate the expedient return and replacement of the product. 3. A mailing label is enclosed for your convenience to return the affected product. Please mark the outside package as "RECALLED PRODUCT" and include the RCL number. All products should be retumed to the following shipping address that is in the letter.
Quantity in commerce
56,694 units
Distribution
Nationwide distribution: USA
Product codes and lots
Catalog No. 300316A (Lot Nos. - NGVJ1108, NGVJ1712, NGVJ1713, NGVJ1105, NGVJ1714) Catalog No. 898316 (Lot Nos. NGVJ1614, NGVJ1615, NGVJ1616, NGVJ1617) Catalog No. 897414 (Lot No. NGVJ1705) Catalog No. 907618 (Lot No. NGVJ1757) Catalog No. 907600 (Lot No. NGVJ1764)

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2327-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2327-2012Class IITerminatedVoluntary: Firm initiated