Z-2227-2012Class III

Access Total T3 Reagent, PN: 33830 The Access Total T3 assay is used for the quantitative determination of...

Siemens Healthcare Diagnostics Inc. / initiated 2011-10-11 / Terminated / root cause: design

Beckman Coulter Inc. began a recall of Access Total T3 Reagent, PN: 33830 The Access Total T3 assay is used for the quantitative determination of triiodothyronine (T3) levels in human serum and... on . FDA classified it as Class III and gives the reason as "A recall was initiated because Beckman Coulter has identified a negative bias in test results when comparing elevated Total T3 patient samples diluted with Sample Diluent A to samples diluted with Access Total T3...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2227-2012
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Access Total T3 Reagent, PN: 33830 The Access Total T3 assay is used for the quantitative determination of triiodothyronine (T3) levels in human serum and plasma in the assessment of thyroid function and diagnosis of thyroid disorders when T3 levels are between 0.1 and 8.0 ng/mL.
Product code
CDP Radioimmunoassay, Total Triiodothyronine
Reason for recall
A recall was initiated because Beckman Coulter has identified a negative bias in test results when comparing elevated Total T3 patient samples diluted with Sample Diluent A to samples diluted with Access Total T3 Calibrator S0.
Root cause tag
design
FDA root cause
Device Design
Action
Beckman Coulter sent a Product Correction letter dated October 11, 2011, to all affected customers. The PCA letter informed the customers of the problem identified and the actions to be taken. Customers were informed to stop SDA usage as an alternative dilutent and that the S0 dilutent is to be used for manual dilution of the patient sample for Total T3. They will also be informed to no longer use the On Board Dilution (OBD) option for Total T3. Customers were instructed to complete and return the enclosed response form within 10 days. Customers with questions regarding this notification were instructed to contact Customer Support Center at http://www.beckmancoulter.com/customersupport/support or call (800) 854-3633 in the US and Canada. Outside of the US and Canada, customers were instructed to contact their local Beckman Coulter Representative. For questions regarding this recall call 714-961-4941.
Quantity in commerce
117,385 units total (33,918 units in US)
Distribution
Worldwide Distribution - USA (nationwide), Canada, and internationally.
Product codes and lots
All Lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2227-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2227-2012Class IIITerminatedVoluntary: Firm initiated