The Brilliance 6, 10, 16, 40, 64 and CT Big Bore are Computed Tomography X-Ray Systems These devices may...
Philips Medical Systems (Cleveland) Inc began a recall of The Brilliance 6, 10, 16, 40, 64 and CT Big Bore are Computed Tomography X-Ray Systems These devices may include signal analysis and display equipment... on . FDA classified it as Class II and gives the reason as "The patient support may move in an unintended manner if the foot switch cover becomes damaged and impinges on the foot switch". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2245-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- The Brilliance 6, 10, 16, 40, 64 and CT Big Bore are Computed Tomography X-Ray Systems These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories. Product Usage: The Brilliance 6, 10, 16, 40, 64 and CT Big Bore are Computed Tomography X-Ray Systems that are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- The patient support may move in an unintended manner if the foot switch cover becomes damaged and impinges on the foot switch.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Philips Medical Systems (Cleveland), Inc. sent an Urgent - Medical Device Correction letter dated August 1, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taked. The letter contained important information for the continued safe and proper use of your equipment. The letter stated that, a Field Service Engineer will visit each affected site to install the hardware upgrade. For further information or support concerning this issue contact your local Philips representative or local Philips Health office. For North America and Canada contact Customer Care Solution Center (1-800-722-9377), option 5: Enter Site ID or Follow the prompts).
- Quantity in commerce
- 126 units
- Distribution
- Worldwide Distribution - US Nationwide including the states of: AL, AR, AZ, CA, DC, FL, GA, KY, MD, MI, MN, MS, ND, NM, NY, OH, OK, SC, TN, TX, VT, WA, and WI and the countries of: Austria, China, Denmark, France, Germany, Iceland, Japan, Netherlands, South Africa, Sweden, Switzerland, United Kingdom. There were two government consignees. There were no Canadian consignees.
- Product codes and lots
- System Code # 728243, Serial #s: 7392, 7394, 7399, 7405, 7406, 7429, 7433, 7440, 7457, 7465, 7469, 7479, 7481, 7487, 7488, 7491, 7502, 7506, 7507, 7517, 7518, 7520, 7521, 7532, 7539, 7545, 7547, and 7556. System Code # 728244, Serial #s: 7190, 7195, 7318, 7383, 7427, 7430, 7510, and 7533. System Code #728256, Serial #s: 3163, 3500, 3500, 5623, 5655, 5655, 5666, 5670, 5715, 5724, 5747, 5773, 5776, 5813, 5836, 5855, 5862, 5868, 5877, 5897, 5911, 5911, 5916, 5935, 5969, and 5997. System Code #728251, Serial #: 4051. System Code # 728246, Serial #s: 5547, 5585, 5791, 5803, 5876, 5974. System Code # 728235, Serial #s: 9233, 9234, 9261, 95022. System Code # 728231, Serial #s: 9867, 9925, 9931, 9980, 9996, 10009, 10011, 10015, 10028, 10036, 10042, 10045, 10050, 10052, 10062, 10063, 10064, 10077, 10079, 10081, 10096, 10099, 10102, 10111, 10114, 10120, 10122, 10133, 10135, 10136, 10137, 10139, 10145, 10148, 10149, 10151, 10153, 10166, 10170, 10172, 10198, 10203, 95109, 95348, 95563.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2245-2012 | Class II | Terminated | Voluntary: Firm initiated |