Z-2339-2012Class II
Tip Cover for 8mm Monopolar Curved Scissors (Hot Shears)Accessory, Disposable, 8mm MCS, Box of 10. The...
Intuitive Surgical, Inc. / initiated 2012-08-09 / Terminated / root cause: manufacturing process
Intuitive Surgical, Inc. began a recall of Tip Cover for 8mm Monopolar Curved Scissors (Hot Shears)Accessory, Disposable, 8mm MCS, Box of 10. The intended use of the Tip Cover Accessory is to provide... on . FDA classified it as Class II and gives the reason as "There is potential for the sterility of the product to be compromised". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2339-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Intuitive Surgical, Inc.
- Product
- Tip Cover for 8mm Monopolar Curved Scissors (Hot Shears)Accessory, Disposable, 8mm MCS, Box of 10. The intended use of the Tip Cover Accessory is to provide dielectric insulation over a section of the Monopolar Curved Scissor instrument so that RF energy is only available at the scissor blades. The Monopolar Curved Scissors are intended for endoscopic manipulation of tissue, including: cutting, blunt and sharp dissection, electrocautery. This accessory is to be used with the Intuitive Surgical Endoscopic Instrument Control Systems (da Vinci, da Vinci S and da Vinci Si Surgical Systems Models IS1200, 12000, ISSOQO).
- Product code
- NAY System, Surgical, Computer Controlled Instrument
- Reason for recall
- There is potential for the sterility of the product to be compromised.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- Intuitive Surgical sent a "MEDICAL DEVICE CORRECTION" letter dated August 9, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. An Acknowledgement Form was included for customers to complete and return via fax to (408) 716-3040. Contact Customer Service at 1-800-876-1310 for questions regarding this recall.
- Quantity in commerce
- 11,121 Boxes
- Distribution
- Worldwide Distribution-USA (nationwide) including Puerto Rico and the countries of Australia, Canada, and Lebanon
- Product codes and lots
- Model No. 400180-12 Lot No: M10120601 M10120430 M10120329 M10120312 M10111214 M10120516 M10120426 M10120321 M10120208 M10111208 M10120510 M10120419 M10120317 M10120207 M10111201 M10120507 M10120413 M10120315 M10120130 M10111028
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2339-2012 | Class II | Terminated | Voluntary: Firm initiated |