Z-2265-2012Class II
Canon Solid State X-ray Imager (Flat Panel/Digital Imager), Digital Radiography, Models CXDI-55C, CXDI-55G...
Canon Inc. / initiated 2010-11-01 / Terminated / root cause: manufacturing process
Canon Inc. began a recall of Canon Solid State X-ray Imager (Flat Panel/Digital Imager), Digital Radiography, Models CXDI-55C, CXDI-55G and CXDI-70C. These devices are intended to replace... on . FDA classified it as Class II and gives the reason as "Screws holding the 55REF circuit board are loosening and roaming inside of the sealed sensor". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2265-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Canon Inc.
- Product
- Canon Solid State X-ray Imager (Flat Panel/Digital Imager), Digital Radiography, Models CXDI-55C, CXDI-55G and CXDI-70C. These devices are intended to replace radiographic film/screen systems in all general purpose diagnostic procedures.
- Product code
- MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)
- Reason for recall
- Screws holding the 55REF circuit board are loosening and roaming inside of the sealed sensor.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Canon Inc. notified sales companies including Canon USA on 9/15/2010 with Service Bulletin 10-017, and on 8/25/2011 with Service Bulletin 11-011, and on 1/20/2011 with Service Bulletin 11-001. Sales representatives were to visit customers and check and replace the screws.
- Quantity in commerce
- 468 units
- Distribution
- Nationwide Distribution (USA)
- Product codes and lots
- Catalog numbers 3930B002AA, 3928B002AA, and 4066B002AA.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2265-2012 | Class II | Terminated | Voluntary: Firm initiated |