Z-0030-2013Class II

Emit(R) 2000 Phenobarbital Assay (OSR4D229) and Syva(R) Emit(R) Phenobarbital Assay (4D019UL)

Siemens Healthcare Diagnostics Inc. / initiated 2012-07-30 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of Emit(R) 2000 Phenobarbital Assay (OSR4D229) and Syva(R) Emit(R) Phenobarbital Assay (4D019UL) Product Usage: Enzyme n Immunoassay, Phenobarbital on . FDA classified it as Class II and gives the reason as "Firm has confirmed highter imprecision with the recalled lots on Beckman Coulter AU(R) Systems. Also, if the Syva(R) Emit(R) 2000 Phenobartbital Assay (4D019UL) lots D4, D5, D6, and/or D7 as an application on AU(R)...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0030-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Emit(R) 2000 Phenobarbital Assay (OSR4D229) and Syva(R) Emit(R) Phenobarbital Assay (4D019UL) Product Usage: Enzyme n Immunoassay, Phenobarbital
Product code
DLZ Enzyme Immunoassay, Phenobarbital
Reason for recall
Firm has confirmed highter imprecision with the recalled lots on Beckman Coulter AU(R) Systems. Also, if the Syva(R) Emit(R) 2000 Phenobartbital Assay (4D019UL) lots D4, D5, D6, and/or D7 as an application on AU(R) Systems is used, the issue may be experienced. Firm has found that within-run imprecison of the recalled lots when run on an AU(R) System may be higher that what would normally be obs
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Siemens sent an Urgent Field Safety notice letter dated July 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately discontinue use and discard any remaining inventory of the affected product. Customers were asked to share the content of the letter with laboratory director regarding the need to review previous Phenobarbital results, conduct patient follow up and/or repeat testing. For a no charge replacement of remaining inventory customers were instructed to contact Siemens Customer Service Center at 888-588-3916. For questions contact Siemens Technical Solutions Center at 800-227-8994.
Quantity in commerce
79,744 cartons
Distribution
United States Nationwide Distribution in the states of: CA, FL, KS, MA, MN, MS, NM, NY, NC, OR, PA, TN, TX, UT, VA, and WA.
Product codes and lots
Emit(R) 2000 Phenobarbital Assay (OSR4D229), lots 1151, 1167, 1191, and/or 1212 and Syva(R) Emit(R) Phenobarbital Assay (4D019UL), lots D4, D5, D6, and D7

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0030-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0030-2013Class IITerminatedVoluntary: Firm initiated