Z-0408-2013Class II

Dimension Vista(R) Systems Vista(R) Theophylline THEO Flex(R) Regent Cartridge (K4071) The THEO method is an...

Siemens Healthcare Diagnostics Inc. / initiated 2012-08-09 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista(R) Systems Vista(R) Theophylline THEO Flex(R) Regent Cartridge (K4071) The THEO method is an in vitro diagnostic test for the quantitative... on . FDA classified it as Class II and gives the reason as "Firm has received complaints from customers of "Abnormal Reaction" errors occurring on calibrations, QC and patient samples. The Abnormal Reaction flag is being triggered by an internal check in the method for atypical...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0408-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Dimension Vista(R) Systems Vista(R) Theophylline THEO Flex(R) Regent Cartridge (K4071) The THEO method is an in vitro diagnostic test for the quantitative measurement of Theophylline in human serum and plasma on the Dimension Vista(R) system.
Product code
KLS Enzyme Immunoassay, Theophylline
Reason for recall
Firm has received complaints from customers of "Abnormal Reaction" errors occurring on calibrations, QC and patient samples. The Abnormal Reaction flag is being triggered by an internal check in the method for atypical particle reagent aggregation.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Siemens Healthcare Diagnostic sent a Urgent Field Safety notice dated July, 20. 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Please immediately discontinue use and discard any remaining inventory of Dimension Vista Theophylline Flex reagent cartridge (THEO, Catalog # K4071) lot 12135AE and/or 12171BB. Please complete the attached form and fax it to (302)631-8467 to indicate that you have received this information and to facilitate your additional no charge replacement. We ask that you forward this notification to anyone to whom you may have distributed this lot. If you have technical questions or concerns, please contact the Siemens Technical Solutions Center at 800-441-9250 for further assistance. For questions concerning the status of your no charge replacement order please contact your Siemens Customer Service Center at 888-588-3916. We apologize for the inconvenience that this situation has caused. Thank you for your patience and continued support.
Quantity in commerce
959 cartons
Distribution
USA (nationwide) including the states of AL, AZ, CA, CO, DC, DE, FL, GA, IL, IN, KY, LA, MA, MD, MI, MO, MS, MT, NC, NH, NJ, NY, OH, OR, PA, SC, TN, TX, VA, VT, WA and WI.
Product codes and lots
Lots 12135AE and 12171BB

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0408-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0408-2013Class IITerminatedVoluntary: Firm initiated