Z-2333-2012Class II

Aesculap S4 Element 4.5mm Polyaxial Screws (ST240T, ST241T, ST242T, ST243T, ST244T, and ST245T). The S4...

Aesculap Inc / initiated 2012-07-25 / Terminated / root cause: manufacturing process

Aesculap Implant Systems LLC began a recall of Aesculap S4 Element 4.5mm Polyaxial Screws (ST240T, ST241T, ST242T, ST243T, ST244T, and ST245T). The S4 Spinal System is intended for anterior/anterolateral... on . FDA classified it as Class II and gives the reason as "There is a possibility that the screw channel that holds the rod in place may be out of tolerance. When the depth of the channel is not deep enough, the rod may not be able to provide ample compression on the insert...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2333-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Aesculap Inc
Product
Aesculap S4 Element 4.5mm Polyaxial Screws (ST240T, ST241T, ST242T, ST243T, ST244T, and ST245T). The S4 Spinal System is intended for anterior/anterolateral and posterior, non-cervical pedicle and non-pedicle fixation. Fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft or allograft. The device is indicated for treatment of the following acute and chronic instabilities or deformities: 1 - degenerative disc disease, 2 - spondylolisthesis, 3 - trauma, 4 - spinal stenosis, 5 - deformities or curvatures, 6 - tumor, 7 - pseudoarthrosis, and 8 - failed previous fusion.
Product code
NKB Thoracolumbosacral Pedicle Screw System
Reason for recall
There is a possibility that the screw channel that holds the rod in place may be out of tolerance. When the depth of the channel is not deep enough, the rod may not be able to provide ample compression on the insert preventing the screw head from locking down to the bone screw.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Aesculap sent an "IMPORTANT PRODUCT REMOVAL NOTIFICATION" letter dated July 25, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. A Distribution Inventory Sheet was attached for customers to complete and return with the affected product. Contact the firm at (610) 984-9275 for questions regarding this recall.
Quantity in commerce
5
Distribution
Nationwide Distribution-including the states of IL, MN, NV, OK, and TX.
Product codes and lots
All distributed lots and serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2333-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2333-2012Class IITerminatedVoluntary: Firm initiated