Z-2377-2012Class II

Capsule Neuron Docking Station, a component in the DataCaptor Connectivity System Catalog Number DC-NU-DS-CK...

Philips North America LLC / initiated 2012-08-10 / Terminated / root cause: design

Capsule Tech Inc. began a recall of Capsule Neuron Docking Station, a component in the DataCaptor Connectivity System Catalog Number DC-NU-DS-CK The Capsule Neuron Docking Station is a component... on . FDA classified it as Class II and gives the reason as "Capsule Neuron Docking Station may experience a lost connection, resulting in loss of data or loss of power to the Caspsule Neuron". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2377-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
Capsule Neuron Docking Station, a component in the DataCaptor Connectivity System Catalog Number DC-NU-DS-CK The Capsule Neuron Docking Station is a component in the DataCaptor Connectivity System. The Capsule Neuron Docking Station provides connection for the Capsule Neuron UMPC which is a bedside device that enables automatic collection of vital signs data that can connect to the hospital network through a wired Ethernet connection or via a standard 802.11 b/g/n wireless network.
Product code
MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)
Reason for recall
Capsule Neuron Docking Station may experience a lost connection, resulting in loss of data or loss of power to the Caspsule Neuron
Root cause tag
design
FDA root cause
Device Design
Action
The firm, Capsule Tech, sent a recall letter dated August 10,2012 to its customers. The letter described the product, problem and actions to be taken. Capsule is sending field correction kits to customers with instructions. Capsule's Professional Services team will contact the customers to schedule shipment of their kit. Final shipments of the field correction kits will be completed by Oct 2012. If you have any questions or concerns, contact your professional services representative and/or call in US: 978-482-2340 or France: +33 153-34-14-00.
Quantity in commerce
12,448 units
Distribution
Worldwide distribution: USA (nationwide) and countries of: Belgium, Canada, Denmark, France, Germany, Netherlands and UK.
Product codes and lots
Serial Numbers TS9801371 -TS9801650; TS9900670 - TS9900679; TS9B00413 - TS9B00722; TS9B01753 - TS9B02041; TSA301587 - TSA302186; TSA600246 - TSA600845 TSA801222 - TSA801821; TSAB00031 - TSAB00629; TSB300410 - TSB300912; TSB302698 - TSB303201; TSB401570 - TSB402012; TSB500633 - TSB501136; TSB502359 - TSB502861; TSB600767 - TSB601270; TSB700368 - TSB700871; TSB801147 - TSB803555; TSB902081 - TSB902585; TSBA03569 - TSBA03595; TSBB00027 - TSBB00537; TSBC00200 - TSBC01321; TSC100319 - TSC101318; TSC101972 - TSC103113; TSC301294 - TSC301299; TSC301389 - TSC303425; TSC401639 - TSC401995; TSC402020 - TSC402087; TSC402136 - TSC402346

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2377-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2377-2012Class IITerminatedVoluntary: Firm initiated