Z-2398-2012Class II

The Bio-Rad MONOLISA Anti-HBs EIA is for In Vitro Diagnostic Use and it has the catalog number 25220. It is a...

Bio-Rad Laboratories Inc. / initiated 2012-07-27 / Terminated / root cause: packaging

Bio-Rad Laboratories began a recall of The Bio-Rad MONOLISA Anti-HBs EIA is for In Vitro Diagnostic Use and it has the catalog number 25220. It is a qualitative and quantitative enzyme immunoassay... on . FDA classified it as Class II and gives the reason as "Prolonged exposure of the MONOLISA Anti-HBs EIA Conjugate Diluent component to light may result in elevated Optical Densities (OD) in some samples such that a negative sample could elevate to a grey zone or low positive...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2398-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Bio-Rad Laboratories Inc.
Product
The Bio-Rad MONOLISA Anti-HBs EIA is for In Vitro Diagnostic Use and it has the catalog number 25220. It is a qualitative and quantitative enzyme immunoassay for the detection of antibody for hepatitis B surface antigen in human serum and EDTA or citrated plasma. The Bio-Rad MONOLISA Anti-HBs EIA is a qualitative and quantitative enzyme immunoassay for the detection of antibody for hepatitis B surface antigen in human serum and EDTA or citrated plasma. The assay results may be used as an aid in the determination of susceptibility to hepatitis B virus (HBV) infection in individuals prior to or following HBV vaccination or where vaccination status is unknown. Assay results may be used with other HBV serological markers for the laboratory diagnosis of HBV disease associated with HBV infection.
Product code
LOM Test, Hepatitis B (B Antigen, B Surface Antigen, Be Antigen)
Reason for recall
Prolonged exposure of the MONOLISA Anti-HBs EIA Conjugate Diluent component to light may result in elevated Optical Densities (OD) in some samples such that a negative sample could elevate to a grey zone or low positive result.
Root cause tag
packaging
FDA root cause
Packaging
Action
The firm, Bio-Rad Laboratories, Inc., sent an "Important Product Information Bio-Rad MONOLISA Anti-HBs EIA Catalog #25220", letter dated 27 July 2012, to their domestic consignees/customers via FedEx. The International subsidiaries will be notified by e-mail. The consignees/customers were informed of the important storage and handling information regarding the MONOLISA Anti-HBs EIA Conjugate Diluent component . The consignees/customers were instructed to follow the storage and handling instructions as stated in the Package Insert, and compare their laboratories practices to the storage and handling instructions stated by Bio-Rad. Customers who have questions can call Bio Rad Technical Support at 1-800-224-6723, option 2, then option 3.
Quantity in commerce
7,400 (approximately)- All lots since distribution date May 9, 2007
Distribution
Worldwide distribution: USA (nationwide) and countries of: Canada, South Africa, and Zimbabwe.
Product codes and lots
All lots distributed from May 5, 2007

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2398-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2398-2012Class IITerminatedVoluntary: Firm initiated