Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2379-2012Class II

HALO Flex Energy Generators Manufactured by BRRX Medical (now Covidien) 540 Oakmead Parkway Sunnyvale, CA...

Medtronic, Inc. / initiated 2012-08-03 / Terminated / root cause: software

Covidien LLC began a recall of HALO Flex Energy Generators Manufactured by BRRX Medical (now Covidien) 540 Oakmead Parkway Sunnyvale, CA 95085. The HALO Flex Energy Generator is indicated... on . FDA classified it as Class II and gives the reason as "Potential failure of HALO Energy Generator to enter "stand by" mode when initially powered on; and therefore can not be used for patient care". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2379-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
HALO Flex Energy Generators Manufactured by BRRX Medical (now Covidien) 540 Oakmead Parkway Sunnyvale, CA 95085. The HALO Flex Energy Generator is indicated for use for the coagulation of soft tissue. The HALO Flex System is indicated for use in the coagulation of bleeding and non-bleeding sites in the gastrointestinal tract including but not limited to, the esophagus, including Barrett's Esophagus.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Potential failure of HALO Energy Generator to enter "stand by" mode when initially powered on; and therefore can not be used for patient care.
Root cause tag
software
FDA root cause
Software design
Action
COVIDIEN sent an Urgent Medical Device Correction letter dated August 3, 2012, to all affected customers via Federal Express. Additionally, Field Service Representatives will be contacting their accounts to notify them of the correction.. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to take the following steps prior to initiating patient treatment and specifically, prior to administering conscious sedation and performing upper endoscopy: 1. Power-on the HALO Flex Energy Generator and confirm that the generator passes self-test, enters standby mode and displays "Ready, Connect Catheter." 2. Confirm that the HALO Flex Energy Generator is functioning properly at power-up and enters the standby mode then the procedure can begin. 3. Do not use the HALO Flex Energy Generator if the generator does not pass the power on self-test. In this situation, please call Covidien, GI Solutions customer service at (888) 662-2779 or your local territory manager (sales representative) so that the firm can provide them with a replacemenent generator. 4. Customers were asked to reply to Covidien, GI Solutions (formerly BARRX Medical) by signing and returning the attached Product Correction Form via fax to the number provided on the form. Customers with questions were instructed to call 888-662-2779. For questions regarding this recall call 408-328-7300.
Quantity in commerce
617 generators
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia, Austria, Belgium, Canada, Chile, Colombia, Czech Republic, Denmark, France, Germany, Greece, Hungary, India, Israel, Jordan, Netherlands, Norway, Poland, Portugal, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, Taiwan, Venezuela and UK.
Product codes and lots
Model numbers 1190A-115A and 1190A-230A; all serial numbers.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2379-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2379-2012Class IITerminatedVoluntary: Firm initiated