Z-2409-2012Class III

ADVIA Centaur Prolactin Assay: ADVIA Centaur Calibrator B Catalog Number. Prolactin:110757 and 110758...

Siemens Healthcare Diagnostics Inc. / initiated 2012-08-07 / Terminated / root cause: manufacturing process

Siemens Healthcare Diagnostics, Inc began a recall of ADVIA Centaur Prolactin Assay: ADVIA Centaur Calibrator B Catalog Number. Prolactin:110757 and 110758: US:Calibrator B 00652707 (6 pack) or 00649625 (2 pack )... on . FDA classified it as Class III and gives the reason as "ADVIA Centaur Prolactin Assay and FSH, Calibrator B fails the analytical sensitivity stated in the IFU for lots below 40". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2409-2012
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
ADVIA Centaur Prolactin Assay: ADVIA Centaur Calibrator B Catalog Number. Prolactin:110757 and 110758: US:Calibrator B 00652707 (6 pack) or 00649625 (2 pack ) OUS: 2 pack REF: 00650577 and Ref: 00654130 (6 pack) Product Usage: ADVIA Centaur CP Prolactin assay: For in vitro diagnostic use in the quantitative determination of prolactin in serum using the ADVIA Centaur System. ADVIA Centaur Calibrator B: For in vitro diagnostic use in calibrating the following assays using ADVIA Centaur Systems: Digoxin, FSH, TotallgE, LH, LH2, Prolactin, Total hCG, and TSH
Product code
CFT Radioimmunoassay, Prolactin (Lactogen)
Reason for recall
ADVIA Centaur Prolactin Assay and FSH, Calibrator B fails the analytical sensitivity stated in the IFU for lots below 40
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Siemens Healthcare issued an Urgent Field Safety Notice letter dated August 2012 by e-mail to all affected Siemens Healthcare Diagnostics Branch offices on August 7, 2012. Customers were instructed to discontinue using the ADVIA Centaur Calibrator B lots 40 and below following their obtaining Calibrator 8 lot 41 or higher. Customers were asked to complete and fax-back the response form attached.
Quantity in commerce
15, 455 kits
Distribution
Worldwide Distribution - US Nationwide and the countries of: Canada, Argentina, Australia, Bangaldesh,Brazil, Chile, China, Columbia, Croatia, Ecuduar, Egypt, Hong Kong, India, Israel,Japan, Korea, Maylasia, New Zealand, Mexico, Peru, Philipines, Pakistan, Saudia Arabia, Singapore, Thailand, Taiwan, Viet Nam, South Africa and UK.
Product codes and lots
US: 2 Pack Ref: 00649625: Lot Numbers/Exp Date: Dec 2, 2012: 78813A30 79982A30 81013A30 82021B30 82767B30 83453C30 85447D30 87319A40 Exp Date:April 9, 2013 85598A40 85708A40 86098A40 87319A40 87979B40 88809B40 89351C40 91750C40 (6 pack) Ref. 00652707: Lot Numbers/Exp Date: Dec 2 2012 79354A30 80802A30 81892B30 82815B30 83741D30 84186A40 Exp. Date April, 8, 2013 85599A40 86191A40 88102B40 89717C40 91383C40 (6 Pack) Ref: 00654254  79714A30 Exp Date Dec 2, 2012 82370A30 84274A30 85512E30 OUS: 2 pack REF: 00650577  78319A29 Exp Dec 2, 2102 78487A29 79024A29 79403A29 79770B29 79981B29 82935B29 85735C29 00654130  Ref: 00654130 (6 pack) 78321A29 Exp Dec 2,2102 79197A29 83314C29

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2409-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2409-2012Class IIITerminatedVoluntary: Firm initiated
Z-2410-2012Class IIITerminatedVoluntary: Firm initiated