Z-2346-2012Class II

Sterile Empty Vial and Injector (30mL PCA vial), Rx only, One/List No. 6021-03, Hospira, Inc., Lake Forest...

Abbott Medical / initiated 2012-08-02 / Terminated / root cause: manufacturing process

Hospira, Inc. began a recall of Sterile Empty Vial and Injector (30mL PCA vial), Rx only, One/List No. 6021-03, Hospira, Inc., Lake Forest, IL 60045. For use with a compatible Hospira PCA... on . FDA classified it as Class II and gives the reason as "Reports of leaking during filling and administration". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2346-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Sterile Empty Vial and Injector (30mL PCA vial), Rx only, One/List No. 6021-03, Hospira, Inc., Lake Forest, IL 60045. For use with a compatible Hospira PCA pump set with injector and a compatible Hospira infusion device.
Product code
MEA Pump, Infusion, Pca
Reason for recall
Reports of leaking during filling and administration.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Hospira sent an Urgent Device Field Correction letter dated August 3, 2012, to all affected customers. The letter identifed the product, the probem, and the action to be taken by the customer. Customers were instructed to first consider the use of Prefilled Morphine PCA Vials wherever possible. In some instances, the treating physician may require an alternative to morphine. Customers may continue using the affected product by following the instructions provided in the Correction Letter. Customers were asked to complete the attached Reply Form and fax it to 1-888-879-8144. For questions regarding this recall call 877-946-7747.
Quantity in commerce
unknown
Distribution
Worldwide Distribution - USA including AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NY, NC, NK, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY and the countries of Canada, Costa Rica and the Virgin Islands.
Product codes and lots
The letter was issued for all product presently in the field and product that will be released. Specific lot numbers were not indicated in the letter.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2346-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2346-2012Class IITerminatedVoluntary: Firm initiated