Z-2337-2012Class III

STERRAD 100S Cassette, Batch # 12A077. The STERRAD 100S and STERRAD NX sterilizations systems are...

Advanced Sterilization Products / initiated 2012-06-19 / Terminated / root cause: labeling and UDI

Advanced Sterilization Products began a recall of STERRAD 100S Cassette, Batch # 12A077. The STERRAD 100S and STERRAD NX sterilizations systems are low-temperature general purpose sterilizers used to sterilize... on . FDA classified it as Class III and gives the reason as "Advanced Sterilization Products (ASP) is recalling (removing) the Sterrad 100S and Sterrad NX System cassettes because the camera used to verify printing on the Sterrad System cassettes was inadvertently disabled. The...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2337-2012
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Advanced Sterilization Products
Product
STERRAD 100S Cassette, Batch # 12A077. The STERRAD 100S and STERRAD NX sterilizations systems are low-temperature general purpose sterilizers used to sterilize heat and moisture sensitive reusable medical devices. The sterilant used is hydrogen peroxide supplied in cassette form as a separate accessory. The STERRAD Cassettes are used in conjunction with the STERRAD Sterilizers, and each cassette model is for use only in the matching sterilizer model.
Product code
MLR Sterilizer, Chemical
Reason for recall
Advanced Sterilization Products (ASP) is recalling (removing) the Sterrad 100S and Sterrad NX System cassettes because the camera used to verify printing on the Sterrad System cassettes was inadvertently disabled. The STERRAD 100S and STERRAD NX System cases were shipped without verification of proper printing.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Advanced Sterilization Products sent an "URGENT MEDICAL DEVICE RECALL (REMOVAL) letter dated July 16, 2012 to all their customers who purchased the STERRAD 100S and STERRAD NX System cassettes. The letter informed the customers of the problem identified and the actions to be taken. For the STERRAD NX System cassettes, there should be no impact to the functionality of these affected cassettes. Although printing may be smeared, illegible, or missing, the packaging has multiple references to the information, and the cassettes should be usable despite any printing error. For the STERRAD 100S System cassettes, there is a chance that the barcode will be unreadable and the STERRAD 100S Sterilizer will not accept the cassette. Customers were instructed to complete and return the business reply card and packing slip that are included in the letter. Customers were instructed to contact Stericycle at (888) 406-9303 for additional labels. Customers with questions about the recall letter were instructed to contact Stericycle at (888) 406-9303.
Quantity in commerce
7,495 units
Distribution
Nationwide Distribution.
Product codes and lots
Product Code: 10113

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2337-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2337-2012Class IIITerminatedVoluntary: Firm initiated
Z-2338-2012Class IIITerminatedVoluntary: Firm initiated