Z-2368-2012Class II

Enduramesh Corpectomy Spacer System

Lucero Medical, LLC / initiated 2012-07-15 / Terminated / root cause: labeling and UDI

Lucero Medical LLC began a recall of Enduramesh Corpectomy Spacer System Product Usage: The Enduramesh is a vertebral body replacement system intended for use in the thoracic and lumbar spine... on . FDA classified it as Class II and gives the reason as "The firm voluntarily recalled the device after they discovered that the sterilization recommendations on the product insert did not pass sterilization validation". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-2368-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Lucero Medical, LLC
Product
Enduramesh Corpectomy Spacer System Product Usage: The Enduramesh is a vertebral body replacement system intended for use in the thoracic and lumbar spine (T1-S1) to replace a collapsed, damaged or unstable vertebral body due to tumor or trauma/fracture. The device is intended for use with either allograft or autograft. The device may be used with a cleared supplemental fixation system based on surgeon discretion and preference.
Product code
MQP Spinal Vertebral Body Replacement Device
Reason for recall
The firm voluntarily recalled the device after they discovered that the sterilization recommendations on the product insert did not pass sterilization validation.
Root cause tag
labeling and UDI
FDA root cause
Error in labeling
Action
Lucero Medical sent a Field Correction Notice letter dated July 15, 2012 to it's sole consignee. The notice identified the affected product., problem and actions to be taken. The customer was instructed to complete the response card and submit back to their local Lucero Medical Representative.
Quantity in commerce
169
Distribution
US Nationwide Distribution
Product codes and lots
Part #-Lot # (s): 762.010L-110738; 762.013L-110738; 762.017L-60611, 90611, and 512171; 762.019L-60611, 90611, 100308, 1261901, and 20121901; 762.021L-60611, 90611, 100308, 110106, 20122101, and 20122102; 762.023L- 90611, 100209, 100218, 00706, 10709, 20122301, and 20122302; 762.025L-90611, 100218, 110106, 20122501, and 20122502; 762.027L-60611, 90611, 100706, and 20122071; 762.029L-90611, 1262901, and 90611; 762.031L-1263101; 762.033L-90611, 91002, and 1263301; 762.041L-90611 and 1264101; 762.043L-90611 and 1264301; 762.045L-90611; 762.047L-90611; 762.049L-90611; 762.051L-90611 and 512511; 762.053L-90611; 762.090L-90611; 762.090L-2.0-100206 and 100342; 762.090L-2.5-100206; 762.159L-110505; 762.237L-91002; 762.243L-91018; 762.270L-91018; 762.290L-91018; 762.731L-91018; 762.735L-91002; 762.767L-91018; and 762.790L-91018.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2368-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2368-2012Class IITerminatedVoluntary: Firm initiated