Enduramesh Corpectomy Spacer System
Lucero Medical LLC began a recall of Enduramesh Corpectomy Spacer System Product Usage: The Enduramesh is a vertebral body replacement system intended for use in the thoracic and lumbar spine... on . FDA classified it as Class II and gives the reason as "The firm voluntarily recalled the device after they discovered that the sterilization recommendations on the product insert did not pass sterilization validation". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2368-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Lucero Medical, LLC
- Product
- Enduramesh Corpectomy Spacer System Product Usage: The Enduramesh is a vertebral body replacement system intended for use in the thoracic and lumbar spine (T1-S1) to replace a collapsed, damaged or unstable vertebral body due to tumor or trauma/fracture. The device is intended for use with either allograft or autograft. The device may be used with a cleared supplemental fixation system based on surgeon discretion and preference.
- Product code
- MQP Spinal Vertebral Body Replacement Device
- Reason for recall
- The firm voluntarily recalled the device after they discovered that the sterilization recommendations on the product insert did not pass sterilization validation.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- Lucero Medical sent a Field Correction Notice letter dated July 15, 2012 to it's sole consignee. The notice identified the affected product., problem and actions to be taken. The customer was instructed to complete the response card and submit back to their local Lucero Medical Representative.
- Quantity in commerce
- 169
- Distribution
- US Nationwide Distribution
- Product codes and lots
- Part #-Lot # (s): 762.010L-110738; 762.013L-110738; 762.017L-60611, 90611, and 512171; 762.019L-60611, 90611, 100308, 1261901, and 20121901; 762.021L-60611, 90611, 100308, 110106, 20122101, and 20122102; 762.023L- 90611, 100209, 100218, 00706, 10709, 20122301, and 20122302; 762.025L-90611, 100218, 110106, 20122501, and 20122502; 762.027L-60611, 90611, 100706, and 20122071; 762.029L-90611, 1262901, and 90611; 762.031L-1263101; 762.033L-90611, 91002, and 1263301; 762.041L-90611 and 1264101; 762.043L-90611 and 1264301; 762.045L-90611; 762.047L-90611; 762.049L-90611; 762.051L-90611 and 512511; 762.053L-90611; 762.090L-90611; 762.090L-2.0-100206 and 100342; 762.090L-2.5-100206; 762.159L-110505; 762.237L-91002; 762.243L-91018; 762.270L-91018; 762.290L-91018; 762.731L-91018; 762.735L-91002; 762.767L-91018; and 762.790L-91018.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2368-2012 | Class II | Terminated | Voluntary: Firm initiated |