Optima XR220amx Mobile general purpose radiographic imaging of the human head and body
GE Healthcare, LLC began a recall of Optima XR220amx Mobile general purpose radiographic imaging of the human head and body on . FDA classified it as Class II and gives the reason as "Software issue: a violation of 21 CFR 1020.31(a). After an image is taken, when the user selects and clears an additionally displayed screen the system inhibit is also cleared, allowing the user to take an exposure...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2255-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- GE Healthcare
- Product
- Optima XR220amx Mobile general purpose radiographic imaging of the human head and body.
- Product code
- IZL System, X-Ray, Mobile
- Reason for recall
- Software issue: a violation of 21 CFR 1020.31(a). After an image is taken, when the user selects and clears an additionally displayed screen the system inhibit is also cleared, allowing the user to take an exposure without the technique factors being displayed.
- Root cause tag
- software
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- The firm will issue Important Electronic Product Radiation Warning letters that identify the affected product along with a description of the defect and related hazards. Users are instructed to ensure that the acquisition mode is always selected prior to attempting an exposure to confirm the techniques selected. The firm will remedy the issue or bring the device into compliance free of charge. A GE Healthcare Service Representative will update the software on the system to address the issue. Field Modification Instructions (FMI) 10865(2) describes the rework plan the GEHC Engineers will make to the imaging system to return it to compliance with the applicable performance standard. It appears to adequately address the problem and is hereby approved. Users are to contact 800-437-1171 in the US, if they have any questions or concerns.
- Quantity in commerce
- 254 total in the US
- Distribution
- Worldwide Distribution -- USA (nationwide) Distribution
- Product codes and lots
- Part Numbers: 5555000-3, 5555000-4, 5555000-5, 5555000-6.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2255-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2256-2012 | Class II | Terminated | Voluntary: Firm initiated |