Z-2256-2012Class II

Optima XR200amx. Mobile general purpose radiographic imaging of the human head and body

GE Healthcare / initiated 2012-07-10 / Terminated / root cause: software

GE Healthcare, LLC began a recall of Optima XR200amx. Mobile general purpose radiographic imaging of the human head and body on . FDA classified it as Class II and gives the reason as "Software issue: a violation of 21 CFR 1020.31(a). After an image is taken, when the user selects and clears an additionally displayed screen the system inhibit is also cleared, allowing the user to take an exposure...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-2256-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
Optima XR200amx. Mobile general purpose radiographic imaging of the human head and body.
Product code
IZL System, X-Ray, Mobile
Reason for recall
Software issue: a violation of 21 CFR 1020.31(a). After an image is taken, when the user selects and clears an additionally displayed screen the system inhibit is also cleared, allowing the user to take an exposure without the technique factors being displayed.
Root cause tag
software
FDA root cause
Radiation Control for Health and Safety Act
Action
The firm will issue Important Electronic Product Radiation Warning letters that identify the affected product along with a description of the defect and related hazards. Users are instructed to ensure that the acquisition mode is always selected prior to attempting an exposure to confirm the techniques selected. The firm will remedy the issue or bring the device into compliance free of charge. A GE Healthcare Service Representative will update the software on the system to address the issue. Field Modification Instructions (FMI) 10865(2) describes the rework plan the GEHC Engineers will make to the imaging system to return it to compliance with the applicable performance standard. It appears to adequately address the problem and is hereby approved. Users are to contact 800-437-1171 in the US, if they have any questions or concerns.
Quantity in commerce
254 total in the US
Distribution
Worldwide Distribution -- USA (nationwide) Distribution
Product codes and lots
Part Numbers: 5555000-3, 5555000-4, 5555000-5, 5555000-6.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2256-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2255-2012Class IITerminatedVoluntary: Firm initiated
Z-2256-2012Class IITerminatedVoluntary: Firm initiated