IDS Urine BETA CrossLaps ELISA, Catalog Number AC-05F1. Type of packaging: Cardboard Box containing plastic...
Immunodiagnostics Systems Ltd began a recall of IDS Urine BETA CrossLaps ELISA, Catalog Number AC-05F1. Type of packaging: Cardboard Box containing plastic bottles. The Urine BETA CrossLaps ELISA is an... on . FDA classified it as Class II and gives the reason as "It has been noted by staff at IDS Inc. that stock of the affected products held at IDS Inc. are labeled with IVD labels although they are not registered for sale as IVDs in the United States. The products should have...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0039-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Immunodiagnostic Systems Ltd.
- Product
- IDS Urine BETA CrossLaps ELISA, Catalog Number AC-05F1. Type of packaging: Cardboard Box containing plastic bottles. The Urine BETA CrossLaps ELISA is an enzyme immunological test for the quantification of degradation products of C-terminal telopeptides of Type-I collagen in human urine. The Urine BETA CrossLaps ELISA assay is intended for in vitro diagnostic use as an indication of human bone resorption as an aid in A. Monitoring bone resorption changes of 1) Anti-resorptive therapies in postmenopausal women: a) Hormone Replacement Therapies (HRT) with hormones and hormone like drugs b) Bisphosphonate therapies B. Predicting skeletal response (Bone Mineral Density) in postmenopausal women undergoing anti-resorptive therapies a) Hormone Replacement Therapies (HRT) with hormones and hormone like drugs b) Bisphosphonate therapies
- Product code
- JMM Column Chromatography & Color Development, Hydroxyproline
- Reason for recall
- It has been noted by staff at IDS Inc. that stock of the affected products held at IDS Inc. are labeled with IVD labels although they are not registered for sale as IVDs in the United States. The products should have been labeled as Research Use Only or Investigational use only dependent on their usage.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling False and Misleading
- Action
- Immunodiagnostics Systems sent Customer notification letters via email and notified customers of the recall. The notification included the reason for recall, product information with codes, lot number, manufacturing/distribution date, and exp date. The notification included instructions to customers: Actions to be taken by the customer: Customers should complete the attached Recall and Acknowledgement Form and take the following actions: 1) Check your stock for the products included within the scope of the recall. Cease use and distribution and quarantine all affected product lots immediately. 2) Complete the enclosed Recall & Acknowledgement Form and immediately fax back to Immunodiagnostic Systems. This will allow us to document your receipt of this letter and the amount of product you have on hand for return. 3) Return any affected product freight collect, along with the original completed Recall & Acknowledgement Form to your local IDS office. Contact information was also provided. For questions regarding this recall call +44 (191) 5195212.
- Quantity in commerce
- 1
- Distribution
- Nationwide Distribution including NY, CA, ME and CT.
- Product codes and lots
- Lot 14541, EXP Nov 2012
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0039-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0040-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0041-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0042-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0043-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0044-2013 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-0045-2013 | Class II | Terminated | Voluntary: Firm initiated |