Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2367-2012Class II

Siemens Mobilett Mira mobile x-ray system

Siemens Medical Solutions USA, Inc. / initiated 2012-04-30 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Siemens Mobilett Mira mobile x-ray system Product Usage: Mobile x-ray system on . FDA classified it as Class II and gives the reason as "During regular product monitoring, firm became aware of a potential issue with the Mobilett Mira system, (serial numbers 1034, 1038, 1041, 1042, 1043, 1050, 1053, and 1054. During normal clinical operation the center...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2367-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Mobilett Mira mobile x-ray system Product Usage: Mobile x-ray system
Product code
IZL System, X-Ray, Mobile
Reason for recall
During regular product monitoring, firm became aware of a potential issue with the Mobilett Mira system, (serial numbers 1034, 1038, 1041, 1042, 1043, 1050, 1053, and 1054. During normal clinical operation the center screw on the driving wheel may become loose resulting in system immobilization. There is no risk of the system tipping over or collapsing on the floor, however the system may tilt t
Root cause tag
design
FDA root cause
Device Design
Action
Siemens has issued an Update Instruction XP018/12S in April 2012 to all affected consignees notifying them of the problem. The update instruction provides a correction of the wheel fixation. For questions call Siemens at 610-219-4834 or 610-448-4634.
Quantity in commerce
9
Distribution
US Nationwide Distribution including the states of: GA, IA, KS, NC and TX.
Product codes and lots
Model number 10273100 - serial numbers 1034, 1038, 1041, 1042, 1043, 1050, 1053, and 1054

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2367-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2367-2012Class IITerminatedVoluntary: Firm initiated