Siemens Mobilett Mira mobile x-ray system
Siemens Medical Solutions USA, Inc began a recall of Siemens Mobilett Mira mobile x-ray system Product Usage: Mobile x-ray system on . FDA classified it as Class II and gives the reason as "During regular product monitoring, firm became aware of a potential issue with the Mobilett Mira system, (serial numbers 1034, 1038, 1041, 1042, 1043, 1050, 1053, and 1054. During normal clinical operation the center...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2367-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Siemens Mobilett Mira mobile x-ray system Product Usage: Mobile x-ray system
- Product code
- IZL System, X-Ray, Mobile
- Reason for recall
- During regular product monitoring, firm became aware of a potential issue with the Mobilett Mira system, (serial numbers 1034, 1038, 1041, 1042, 1043, 1050, 1053, and 1054. During normal clinical operation the center screw on the driving wheel may become loose resulting in system immobilization. There is no risk of the system tipping over or collapsing on the floor, however the system may tilt t
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Siemens has issued an Update Instruction XP018/12S in April 2012 to all affected consignees notifying them of the problem. The update instruction provides a correction of the wheel fixation. For questions call Siemens at 610-219-4834 or 610-448-4634.
- Quantity in commerce
- 9
- Distribution
- US Nationwide Distribution including the states of: GA, IA, KS, NC and TX.
- Product codes and lots
- Model number 10273100 - serial numbers 1034, 1038, 1041, 1042, 1043, 1050, 1053, and 1054
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2367-2012 | Class II | Terminated | Voluntary: Firm initiated |