Integra HRAIM Head Ring Assembly with Intubation Mounts The CRW stereotactic system is used in conjunction...

Integra LifeSciences Corporation / initiated 2012-08-09 / Terminated / root cause: design

Integra LifeSciences Corporation began a recall of Integra HRAIM Head Ring Assembly with Intubation Mounts The CRW stereotactic system is used in conjunction with the HRAIM head ring is a target-centered system... on . FDA classified it as Class II and gives the reason as "T-handle screw is used on a complementary product to the CRW, the Integra HRAIM, which is the CT-compatible intubation Head Ring Assembly. When the T-handle on the HRAIM intubation hoop is tightened, the T-handle could...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2429-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra LifeSciences Corporation
Product
Integra HRAIM Head Ring Assembly with Intubation Mounts The CRW stereotactic system is used in conjunction with the HRAIM head ring is a target-centered system that can be configured to be CT-only or CT/MR compatible. The CRW stereotactic system is a multi-purpose system used for localizing intercranial targets for precisely directing instruments such as: Biopsy forceps RF lesioning electrodes Deep brain electrodes Recording and stimulating electrodes
Product code
HAW Neurological Stereotaxic Instrument
Reason for recall
T-handle screw is used on a complementary product to the CRW, the Integra HRAIM, which is the CT-compatible intubation Head Ring Assembly. When the T-handle on the HRAIM intubation hoop is tightened, the T-handle could stop in a vertical position that prevents the CRWPRECISE and certain models of the CRW-ASL from seating properly on the head ring although it appears to be fully seated
Root cause tag
design
FDA root cause
Device Design
Action
The firm, Integra, sent an "URGENT: VOLUNTARY MEDICAL DEVICE RECALL NOTIFICATION" dated August 10, 2012 to US & Canada consignees/customers via traceable courier service and Non-US & Non-Canada consignees/customers - through written notification delivered either by traceable courier, email, facsimile, or through a directed visit by an Integra Sales Representative. The notification describes the product, problem and actions to be taken. The Integra sales representative will replace the screws according to the attached rework instructions. If for some reason it is not possible to replace the screws during the visit, the sales representative will leave the screws and customers can then replace them by following the attached rework instructions and discard the old screws immediately after replacement. The customers were instructed to complete and return the Product Recall Acknowledgement Form via the sales representatives visit; fax at 1-609-750-7999 or email to: [email protected] and include the number of devices to be identified and replaced . Note:The T -Handle screws that are replaced will be disposed of at the consignee's/customers location. Please feel free to contact our service hotline at 888-772-7378 should you have any additional questions.
Quantity in commerce
188 units
Distribution
Worldwide distribution: USA (Nationwide) and countries of: Argentina, Australia, Belgium, Canada, China, Czec, Denmark, France, Hong Kong, Hungary, Italy, Japan, Malaysia, Philippines, Singapore, Spain, South Africa, and UK.
Product codes and lots
All lot numbers from 2002 to current

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2429-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2428-2012Class IITerminatedVoluntary: Firm initiated
Z-2429-2012Class IITerminatedVoluntary: Firm initiated