The Brilliance CT 16 System
Philips Medical Systems (Cleveland) Inc began a recall of The Brilliance CT 16 System Product Usage: The Brilliance 16 is a Whole Body Computed Tomography X-ray System intended to produce cross-sectional images of the... on . FDA classified it as Class III and gives the reason as "Philips was notified that the system logout in software version 2.3.6 is now longer and may sometimes fail to logout, requiring a forced system restart or hard shutdown with the power button". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2404-2012
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- The Brilliance CT 16 System Product Usage: The Brilliance 16 is a Whole Body Computed Tomography X-ray System intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- Philips was notified that the system logout in software version 2.3.6 is now longer and may sometimes fail to logout, requiring a forced system restart or hard shutdown with the power button.
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- Philips sent an Urgent - Medical Device Correction letter dated August 24, 2012 via certified mail to all US consignees. Philips Key Market representatives distributed the letter internationally. The letter identified the affected product, problem and actions to be taken by customer/user. Philips is implementing a software update to correct the issue free of charge. A Philips representative will contact affect sites to schedule the software installation. For further information or support contact your Philips representative or local Philips Healthcare office.
- Quantity in commerce
- 93 units
- Distribution
- Worldwide Distribution - US Nationwide including the states of: AL, CA, GA, MA, MD, NC, OH, PA, TX, and WA; and to the following countries of: Algeria, Oman, Australia, Brazil, China, Saudi Arabia, United Arab Emirates, and India.
- Product codes and lots
- System Code #, 728246, Serial #, 6462, 6447, 6495, 6514, 6467, 6482, 6473, 6544, 6439, 6453, 6436, 6536, 6448, 6449, 6471, 6498, 6502, 6504, 6452, 6510, 6508, 6438, 6513, 6456, 6505, 6457, 6442, 6454, 6461, 6469, 6485, 6458, 6475, 6487, 6531, 6520, 6490, 6470, 6437, 6497, 6481, 6489, 6488, 6455, 6543, 6499, 6493, 6443, 6463, 6476, 6477, 6478, 6526, 6545, 6546, 6518, 6491, 6450, 6451, 6468, 6464, 6466, 6459, 6472, 6444, 6445, 6529, 6530, 6440, 6532, 6515, 6486, 6507, 6512, 6483, 6524, 6480, 6479, 6509, 6460, 6441, 6465, 6474, 6484, 6492, 6503, 6511, 6517, 6519, 6523, 6525, 6549, 6555
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2404-2012 | Class III | Terminated | Voluntary: Firm initiated |