Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2386-2012Class II

MX 16-slice CT Scanner System, models 728131 and 728132. The MX 16-slice CT Scanner System can be used as a...

Philips North America LLC / initiated 2012-08-24 / Terminated / root cause: manufacturing process

Philips Medical Systems (Cleveland) Inc began a recall of MX 16-slice CT Scanner System, models 728131 and 728132. The MX 16-slice CT Scanner System can be used as a whole body computed tomography X-ray system on . FDA classified it as Class II and gives the reason as "On August 24, 2012 the firm recalled the MX 16-slice after discovering that a bolt which attaches the oil accumulator within the gantry may fail, resulting in the oil accumulator detaching from the tube housing assembly...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2386-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
MX 16-slice CT Scanner System, models 728131 and 728132. The MX 16-slice CT Scanner System can be used as a whole body computed tomography X-ray system.
Product code
JAK System, X-Ray, Tomography, Computed
Reason for recall
On August 24, 2012 the firm recalled the MX 16-slice after discovering that a bolt which attaches the oil accumulator within the gantry may fail, resulting in the oil accumulator detaching from the tube housing assembly within the gantry.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
An "Urgent Medical Device Correction" Field Safety Notice dated August 24, 2012 was sent to all US consignees via certified mail. The problem was described and recommended actions were provided. Philips Key Market representatives distributed the letter internationally. Additionally, a Field Service Engineer will contact each site to schedule time to implement the Field Change Orders. US Customer contact number: 1 800-722-9377.
Quantity in commerce
328 units
Distribution
Worldwide distribution, Nationwide (US) including the states AL, AR, CA, FL, GA, IL, IN, MS, NJ, NV, NY, OH, PA, and VA and the countries of Albania, Argentina, Australia, Belgium, Brazil, Chile, China, Colombia, Dominican Republic, Ecuador, France, Germany, Greece, India, Indonesia, Ireland, Italy, Jordan, Lebanon, Malawi, Mexico, Nepal, Netherlands, Oman, Palestine, Philippines, Poland, Romania, Russia, Saudi Arabia, South Africa, South Korea, Spain, Switzerland, Turkey, Uganda, and Venezuela.
Product codes and lots
System Code #, 728131, Serial #: EP16E090001-EP16E090004, EP16E090006-EP16E090015, EP16E100001- EP16E100004, EP16E100006, EP16E100008- EP16E100063, EP16E100067, EP16E100069- EP16E100113, EP16E110001- EP16E110065, EP16E110067-EP16E110070, EP16E110074-EP16E110075, EP16E110077-EP16E110078, EP16E110080, EP16E110083- EP16E110085, EP16E110088- EP16E110112, EP16E110115- EP16E110119, EP16E120001- EP16E120014, EP16E120016, EP16EB110003- EP16ER100007, EP16ER110003, EP16ER110006, P16C080001, P16C080002, P16C090001, P16C100001-P16C100015, P16C100017, P16C100018, P16C110001- P16C110008, P16C110011-P16C110037, P16C120001-P16C120003, P16C120006, P16C120007, P16E080005, P16E080007, P16E080008, P16E080012, P16E090002-P16E090021, P16E090024-P16E090028, P16E100002, P16EF80010, P16EM08003, and P16EM09002-P16EM09004.  System Code #, 728132, Serial #: EP16E100005, EP16E100007, EP16E110066, EP16ER100003-EP16ER100005, EP16ER100008, EP16ER110004, EP16ER110005, and EP16ER110007-EP16ER110024.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2386-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2386-2012Class IITerminatedVoluntary: Firm initiated