MX 16-slice CT Scanner System, models 728131 and 728132. The MX 16-slice CT Scanner System can be used as a...
Philips Medical Systems (Cleveland) Inc began a recall of MX 16-slice CT Scanner System, models 728131 and 728132. The MX 16-slice CT Scanner System can be used as a whole body computed tomography X-ray system on . FDA classified it as Class II and gives the reason as "On August 24, 2012 the firm recalled the MX 16-slice after discovering that a bolt which attaches the oil accumulator within the gantry may fail, resulting in the oil accumulator detaching from the tube housing assembly...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2386-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Philips North America LLC
- Product
- MX 16-slice CT Scanner System, models 728131 and 728132. The MX 16-slice CT Scanner System can be used as a whole body computed tomography X-ray system.
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- On August 24, 2012 the firm recalled the MX 16-slice after discovering that a bolt which attaches the oil accumulator within the gantry may fail, resulting in the oil accumulator detaching from the tube housing assembly within the gantry.
- Root cause tag
- manufacturing process
- FDA root cause
- Process design
- Action
- An "Urgent Medical Device Correction" Field Safety Notice dated August 24, 2012 was sent to all US consignees via certified mail. The problem was described and recommended actions were provided. Philips Key Market representatives distributed the letter internationally. Additionally, a Field Service Engineer will contact each site to schedule time to implement the Field Change Orders. US Customer contact number: 1 800-722-9377.
- Quantity in commerce
- 328 units
- Distribution
- Worldwide distribution, Nationwide (US) including the states AL, AR, CA, FL, GA, IL, IN, MS, NJ, NV, NY, OH, PA, and VA and the countries of Albania, Argentina, Australia, Belgium, Brazil, Chile, China, Colombia, Dominican Republic, Ecuador, France, Germany, Greece, India, Indonesia, Ireland, Italy, Jordan, Lebanon, Malawi, Mexico, Nepal, Netherlands, Oman, Palestine, Philippines, Poland, Romania, Russia, Saudi Arabia, South Africa, South Korea, Spain, Switzerland, Turkey, Uganda, and Venezuela.
- Product codes and lots
- System Code #, 728131, Serial #: EP16E090001-EP16E090004, EP16E090006-EP16E090015, EP16E100001- EP16E100004, EP16E100006, EP16E100008- EP16E100063, EP16E100067, EP16E100069- EP16E100113, EP16E110001- EP16E110065, EP16E110067-EP16E110070, EP16E110074-EP16E110075, EP16E110077-EP16E110078, EP16E110080, EP16E110083- EP16E110085, EP16E110088- EP16E110112, EP16E110115- EP16E110119, EP16E120001- EP16E120014, EP16E120016, EP16EB110003- EP16ER100007, EP16ER110003, EP16ER110006, P16C080001, P16C080002, P16C090001, P16C100001-P16C100015, P16C100017, P16C100018, P16C110001- P16C110008, P16C110011-P16C110037, P16C120001-P16C120003, P16C120006, P16C120007, P16E080005, P16E080007, P16E080008, P16E080012, P16E090002-P16E090021, P16E090024-P16E090028, P16E100002, P16EF80010, P16EM08003, and P16EM09002-P16EM09004. System Code #, 728132, Serial #: EP16E100005, EP16E100007, EP16E110066, EP16ER100003-EP16ER100005, EP16ER100008, EP16ER110004, EP16ER110005, and EP16ER110007-EP16ER110024.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2386-2012 | Class II | Terminated | Voluntary: Firm initiated |