Z-0023-2013Class II

PREVI Isola System, Catalog numbers 29500 and 29500R, bioMerieux, Inc., Durham, NC. Intended to perform the...

bioMerieux, Inc. / initiated 2012-08-28 / Terminated / root cause: labeling and UDI

Biomerieux Inc began a recall of PREVI Isola System, Catalog numbers 29500 and 29500R, bioMerieux, Inc., Durham, NC. Intended to perform the inoculation and streaking of Prepoured Media (PPM)... on . FDA classified it as Class II and gives the reason as "The instrument may not dispense the specimen onto the pre-poured media resulting in a false negative result". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0023-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
bioMerieux, Inc.
Product
PREVI Isola System, Catalog numbers 29500 and 29500R, bioMerieux, Inc., Durham, NC. Intended to perform the inoculation and streaking of Prepoured Media (PPM) with liquid microbiological samples.
Product code
LXG Equipment, Laboratory, General Purpose, Labeled Or Promoted For A Specific Medical Use
Reason for recall
The instrument may not dispense the specimen onto the pre-poured media resulting in a false negative result.
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
bioMerieux sent an "URGENT PRODUCT CORRECTION NOTICE" letter dated August 28, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. An Acknowledgement Form was included for customers to complete and return. Contact Customer Service at 1-800-682-2666 for questions regarding this notice.
Quantity in commerce
350 systems
Distribution
Worldwide Distribution-USA (nationwide)-including DC and the states of CA, CO, CT, FL, GA, IL, IN, KY, LA, MO, NC, NJ, NY, OH, OR, PA, RI, SC, TX, and WI, and the country of Canada.
Product codes and lots
Serial numbers AS180-00001 to AS180-00376.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0023-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0023-2013Class IITerminatedVoluntary: Firm initiated