Z-0114-2013Class II

GE Healthcare innova 3100-IQ, Mobile X-ray system The innova systems are indicated for use in generating...

GE Healthcare / initiated 2011-12-20 / Terminated

GE Healthcare, LLC began a recall of GE Healthcare innova 3100-IQ, Mobile X-ray system The innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular... on . FDA classified it as Class II and gives the reason as "It was discovered of a potential failure of the spectral filter mechanism inside the collimator of the innova 3100, innova 3100-IQ and innova 3131-IQ systems". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0114-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
GE Healthcare
Product
GE Healthcare innova 3100-IQ, Mobile X-ray system The innova systems are indicated for use in generating fluoroscopic images of human anatomy for vascular angiography, diagnostic and interventional procedures, and optionally , rotational angiography procedures. They are also indicated for generating fluoroscopic images of human anatomy for cardiology, diagnostic, and interventional procedures.
Product code
IZI System, X-Ray, Angiographic
Reason for recall
It was discovered of a potential failure of the spectral filter mechanism inside the collimator of the innova 3100, innova 3100-IQ and innova 3131-IQ systems.
FDA root cause
Radiation Control for Health and Safety Act
Action
GE Healthcare will, without charge, remedy the defect or bring the product into compliance with each applicable Federal standard in accordance with a plan to be approved by the Secretary of Health and Human Services. A GE Healthcare Service Representative will replace the collimator to address the issue. Notification of all dealers and purchasers is to be made within 15 working days of receipt of this letter in the manner specified in 21 CFR 1003.21 and 1003.22. This office and the Food and Drug Administration (FDA) district office coordinator noted below are to be included in the notification. Further questions please call 262-513-4122.
Quantity in commerce
199 installed in the US
Distribution
Nationwide distribution
Product codes and lots
Model number - 5892935G052G

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0114-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0113-2013Class IITerminatedVoluntary: Firm initiated
Z-0114-2013Class IITerminatedVoluntary: Firm initiated
Z-0115-2013Class IITerminatedVoluntary: Firm initiated