Varian brand Clinac, Trilogy, Novalis Tx, Unique, linear accelerators; Model Number: H14, H18, H29...
Varian Medical Systems, Inc. Oncology Systems began a recall of Varian brand Clinac, Trilogy, Novalis Tx, Unique, linear accelerators; Model Number: H14, H18, H29, Reference/FSCA Identifier: CP-08881; Product is... on . FDA classified it as Class II and gives the reason as "Some X-jaw (lower collimator jaw) carriers were made using the incorrect metal alloy, which can cause the jaw carrier to crack. If both carriers on the same X-jaw crack, the jaw would be able to move freely in the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2383-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Varian Medical Systems, Inc.
- Product
- Varian brand Clinac, Trilogy, Novalis Tx, Unique, linear accelerators; Model Number: H14, H18, H29, Reference/FSCA Identifier: CP-08881; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA. 1) The Varian Low Energy Linear Accelerator is indicated for precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment i is indicated. (2) The Varian High Energy Linear Accelerator is indicated for stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
- Product code
- IYE Accelerator, Linear, Medical
- Reason for recall
- Some X-jaw (lower collimator jaw) carriers were made using the incorrect metal alloy, which can cause the jaw carrier to crack. If both carriers on the same X-jaw crack, the jaw would be able to move freely in the closed direction. The position readout interlock circuitry will not detect this jaw position deviation. This may lead to treatment with an under-dose of the intended target volume.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Varian sent an "URGENT MEDICAL DEVICE CORRECTION URGENT FIELD SAFETY NOTICE" dated August 14, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 1-888-827-4265 for questions regarding this recall.
- Quantity in commerce
- 221 Units
- Distribution
- Worldwide Distribution.
- Product codes and lots
- LIST OF SERIAL NUMBERS: H140863 ,H294608 ,H294669 ,H294732 ,H140865 ,H294610 ,H294670 ,H294733 ,H140866 ,H294611 ,H294671 ,H294734 ,H140867 ,H294612 ,H294672 ,H294735 ,H140868 ,H294613 ,H294673 ,H294736 ,H140869 ,H294614 ,H294674 ,H294737 ,H140870 ,H294615 ,H294675 ,H294738 ,H140871 ,H294616 ,H294676 ,H294739 ,H140872 ,H294617 ,H294677 ,H294740 ,H140873 ,H294618 ,H294678 ,H294741 ,H140874 ,H294619 ,H294679 ,H294742 ,H140875 ,H294620 ,H294680 ,H294743 ,H140876 ,H294621 ,H294681 ,H294744 ,H140877 ,H294622 ,H294682 ,H294745 ,H140878 ,H294623 ,H294683 ,H294746 ,H140879 ,H294624 ,H294684 ,H294747 ,H140880 ,H294625 ,H294685 ,H294748 ,H140881 ,H294626 ,H294686 ,H294749 ,H140882 ,H294627 ,H294687 ,H294750 ,H140883 ,H294628 ,H294688 ,H294751 ,H140884 ,H294629 ,H294689 ,H294752 ,H140885 ,H294630 ,H294690 ,H294753 ,H140886 ,H294631 ,H294691 ,H294754 ,H140887 ,H294632 ,H294692 ,H294755 ,H140888 ,H294633 ,H294693 ,H294756 ,H140889 ,H294634 ,H294694 ,H294757 ,H140890 ,H294635 ,H294695 ,H294758 ,H140891 ,H294636 ,H294696 ,H294759 ,H140892 ,H294637 ,H294697 ,H294760 ,H140893 ,H294638 ,H294698 ,H294761 ,H140895 ,H294639 ,H294699 ,H294762 ,H140896 ,H294640 ,H294700 ,H294763 ,H140897 ,H294641 ,H294701 ,H294764 ,H140898 ,H294642 ,H294702 ,H294765 ,H140899 ,H294643 ,H294704 ,H294766 ,H140900 ,H294644 ,H294705 ,H294767 ,H181175 ,H294645 ,H294706 ,H294768 ,H181176 ,H294646 ,H294707 ,H294769 ,H181177 ,H294647 ,H294708 ,H294770 ,H181178 ,H294648 ,H294709 ,H294771 ,H181179 ,H294649 ,H294710 ,H294772 ,H181180 ,H294650 ,H294711 ,H181181 ,H294651 ,H294712 ,H181182 ,H294652 ,H294713 ,H181183 ,H294653 ,H294714 ,H181184 ,H294654 ,H294715 ,H181185 ,H294655 ,H294716 ,H181186 ,H294656 ,H294717 ,H181187 ,H294657 ,H294718 ,H181188 ,H294658 ,H294719 ,H181189 ,H294659 ,H294720 ,H181190 ,H294660 ,H294722 ,H181191 ,H294661 ,H294723 ,H181192 ,H294662 ,H294724 ,H182000 ,H294663 ,H294726 ,H182001 ,H294664 ,H294727 ,H182002 ,H294665 ,H294728 ,H182003 ,H294666 ,H294729 ,H182005 ,H294667 ,H294730 ,H294607 ,H294668 ,H294731
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2383-2012 | Class II | Terminated | Voluntary: Firm initiated |