Siemens Mobilett Mira. Mobile X-Ray System

Siemens Medical Solutions USA, Inc. / initiated 2012-08-21 / Terminated / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Siemens Mobilett Mira. Mobile X-Ray System on . FDA classified it as Class II and gives the reason as "During regular product monitoring, Siemens has become aware of a potential issue with the Mobilett Mira system. During normal clinical operation the switch on the motor drive handle of the unit may become loose and...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-2322-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Siemens Mobilett Mira. Mobile X-Ray System.
Product code
IZL System, X-Ray, Mobile
Reason for recall
During regular product monitoring, Siemens has become aware of a potential issue with the Mobilett Mira system. During normal clinical operation the switch on the motor drive handle of the unit may become loose and slide horizontally to the right and become trapped under the cover. This may cause the system to continuously move. In response, Siemens has issued Update Instruction XP038/12/S to r
Root cause tag
design
FDA root cause
Device Design
Action
Siemens initiated their correction of the MOBILETT Mira on August 21, 2012 by sending an "UPDATE INSTRUCTIONS XP038/12/S to all affected customers. The Instruction identifies the product, problem, and work steps for customers to follow. A XP038/12/S Completion Protocol letter was included for customers to complete and file with the instruction once the update is completed.
Quantity in commerce
20
Distribution
Nationwide Distribution- including the states of GA, ID, IL, IA, KS, NC, OH, SD, TX, and WI.
Product codes and lots
Model number 10273100 -- serial numbers 1034, 1038, 1041, 1042, 1043, 1053, 1057, 1070, 1079, 1081, 1088, 1098, 1099, 1102, 1103, 1105, 1106, 1107, 1108, and 1112.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2322-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2322-2012Class IITerminatedVoluntary: Firm initiated