Z-0069-2013Class I

Symbiq One Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids; Hospira, Inc...

Abbott Medical / initiated 2012-08-29 / Terminated / root cause: design

Hospira Inc. began a recall of Symbiq One Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids; Hospira, Inc., Lake Forest, IL 60045 USA, Made in Costa Rica... on . FDA classified it as Class I and gives the reason as "The Symbiq pump touchscreen may not respond to user selection, may experience a delayed response, or may register a different value from the value selected by the user. Failure of the touchscreen to respond to user...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0069-2013
Classification
Class I
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Symbiq One Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids; Hospira, Inc., Lake Forest, IL 60045 USA, Made in Costa Rica; List Number 16026-01 Device is a general purpose infusion pump designed to deliver fluids, solutions, medications, agents, nutritionals, electrolytes, blood and blood products for parenteral, enteral, intravenous, intra-arterial, subcutaneous, epidural
Product code
FRN Pump, Infusion
Reason for recall
The Symbiq pump touchscreen may not respond to user selection, may experience a delayed response, or may register a different value from the value selected by the user. Failure of the touchscreen to respond to user input could result in a delay in therapy or result in over delivery or under delivery of medication if the user does not confirm the programmed values on the pump's confirmation screen prior to starting the infusion.
Root cause tag
design
FDA root cause
Device Design
Action
Hospira sent an Urgent Device Correction letter dated August 29, 2012, to all affected customers. Customers have been informed of reports that the Symbiq pump touchscreen may not respond to user selection, may experience a delayed response, or may register a different value from the value selected by the user. Failure of the touchscreen to respond to user input could result in a delay in therapy or result in over delivery or under delivery of medication if the user does not confirm the programmed values on the pump's confirmation screen prior to starting the infusion. The users were instructed to complete and fax the enclosed reply card to 1-888-943-5180 or email it to [email protected] indicating the number of infusers at the facility and if they notified their users throughout the facility. Wholesalers were asked to notify their customers. If they have further distributed the affected product, they should notify their accounts and ask them to contact Stericycle at 1-877-272-6141 to receive a reply form. For questions regarding this recall call 224-212-2000.
Quantity in commerce
19,614 units
Distribution
Worldwide Distribution - USA (nationwide) Puerto Rico and Internationally to Canada
Product codes and lots
List Number 16026-01, all serial numbers

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0069-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0069-2013Class ITerminatedVoluntary: Firm initiated
Z-0070-2013Class ITerminatedVoluntary: Firm initiated