00780303201 MIS 12/14 Slotted Femoral Head Provisional -3.5X32 00780303202 MIS 12/14 Slotted Femoral Head...
Zimmer, Inc. began a recall of 00780303201 MIS 12/14 Slotted Femoral Head Provisional -3.5X32 00780303202 MIS 12/14 Slotted Femoral Head Provisional +0X32 00780303203 MIS 12/14 Slotted... on . FDA classified it as Class II and gives the reason as "Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2482-2012
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- 00780303201 MIS 12/14 Slotted Femoral Head Provisional -3.5X32 00780303202 MIS 12/14 Slotted Femoral Head Provisional +0X32 00780303203 MIS 12/14 Slotted Femoral Head Provisional +3.5X32 00780303204 MIS 12/14 Slotted Femoral Head Provisional +7X32 00780303205 MIS 12/14 Slotted Femoral Head Provisional +10.5X32 00780303214 MIS 12/14 Slotted Femoral Head Provisional +7X32 The provisional heads replicate final femoral implant geometry and are used to trial range of motion, joint stability, and determine appropriate femoral head size and offset during total hip arthroplasty.
- Product code
- HWT Template
- Reason for recall
- Complaints have been received reporting sub-optimal fit between provisional femoral heads (VerSys and MIS) and the mating components (ie stems and provisional femoral necks), in some cases alleging the fit was too loose and in other cases, alleging the fit was too tight. Intraoperative dislodging of the provisional head and the provisional head being left in-vivo after separating from mating comp
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Zimmer, Inc. sent an E-mail letter dated April 21, 2011, to all affected customers in the US. A June 2011 E-mail was sent to all customers outside the US. The letters requested Inventory & Operations Managers to check product for proper fit /function, complete the return form with worn provisional heads to Zimmer Inc, Attention Product Service, 1777 West Center St. Warsaw, IN 46580. For questions regarding this recall call 574-372-4753.
- Quantity in commerce
- 231,641
- Distribution
- Worldwide Distribution - USA (nationwide) and internationally to Canada, Mexico, South America, Europe, Puerto Rico, China, and Korea.
- Product codes and lots
- All lots including lot 78939000 (1/7/2002) to present
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2473-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2474-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2475-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2476-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2477-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2478-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2479-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2480-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2481-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2482-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2483-2012 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-2484-2012 | Class II | Terminated | Voluntary: Firm initiated |