AC Powered adjustable hospital bed, Prime Bed TL500, Distributed by: Primus Medical LLC, 8401 Southern Blvd...
Primus Medical LLC began a recall of AC Powered adjustable hospital bed, Prime Bed TL500, Distributed by: Primus Medical LLC, 8401 Southern Blvd, Boardman, OH 44512. Usage: Hospital Bed on . FDA classified it as Class II and gives the reason as "The "Retaining washers & Clevis Pins" provided with the AC -powered adjustable hospital beds may fail under certain circumstances. The pins at the top of the leg were disengaging from the channel. This may cause the bed...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0270-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Primus Medical LLC
- Product
- AC Powered adjustable hospital bed, Prime Bed TL500, Distributed by: Primus Medical LLC, 8401 Southern Blvd, Boardman, OH 44512. Usage: Hospital Bed.
- Product code
- FNL Bed, Ac-Powered Adjustable Hospital
- Reason for recall
- The "Retaining washers & Clevis Pins" provided with the AC -powered adjustable hospital beds may fail under certain circumstances. The pins at the top of the leg were disengaging from the channel. This may cause the bed frame to come off tracks or collapse.
- Root cause tag
- manufacturing process
- FDA root cause
- Equipment maintenance
- Action
- The firm, Primus Medical LLC, sent an "Important Notification Relating to Primus Medical Product TL500" letter dated March 29, 2011 to their customers. The letter described the product, problem and actions to be taken. The customers were instructed to replace the "Retaining washers & Clevis Pins" with the "Clevis Pins & Bow Tie Cotter Pins" in the TL500-001 Service Pack provided with the letter in order to avoid any potential failure and/or bed collapse and to return the" Retaining washers & Clevis Pins" after the replacement procedure is performed. If you have any questions, or need assistance in performing the replacement, call 330-965-9681.
- Quantity in commerce
- 342 Beds
- Distribution
- Nationwide distribution: USA including states of: AZ, IL, KY, MD, MI, OH, OR, PA, TX and VA.
- Product codes and lots
- Model #TL500; Serial Number: 001, 002, 003, 004, 005, 006, 007, 010, 013, 014, 015, 016, 017, 018, 021, 022, 023, 024, 025, 026, 027, 028, 029, 030, 031, 032, 033, 034, 035, 039, 040, 041, 042, 043, 044, 045, 046, 047, 048, 049, 050, 052, 053, 054, 055, 056, 057, 058, 059, 060, 061, 062, 063, 064, 065, 070, 075, 081, 09350014, 09350015, 09350016, 09350017, 09350018, 09350021, 09350023, 09350025, 09350026, 09350027, 09350028, 09350029, 09350030, 09350031, 09350032, 09350033, 09350035, 09350038, 09350040, 09350041, 09350042, 09350045, 09350048, 09350049, 09350054, 09350055, 09350056, 09350057, 09350058, 09350059, 09350060, 09350062, 09350063, 09350064, 09350065, 09350066, 09350067, 09350068, 09350069, 09350070, 09350072, 09350074, 09350082, 09350085, 09350087, 09350088, 09350089, 09350090, 09350091, 09350092, 09430001, 09430002, 09430003, 09430004, 09430005, 09430006, 09430007, 09430008, 09430010, 09430013, 09430014, 09430015, 09430017, 09430019, 09430020, 09430021, 09430022, 09430024, 09430026, 09430027, 09430028, 09430029, 09430030, 09430032, 09430033, 09430034, 09430036, 09430037, 09430038, 09430039, 09430040, 09430041, 09430042, 09430043, 09430044, 09430045, 09430046, 09430047, 09430049, 09430050, 09430051, 09430052, 09430053, 09430054, 09430056, 09430057, 09430058, 09430059, 09430060, 09430061, 09430062, 09430063, 09430065, 09430066, 09463067, 09430068, 09430069, 09430070, 09430071, 09430072, 09430074, 09430075, 09430076, 09430077, 09430078, 09430079, 09430080, 09430081, 09430083, 09430085, 09430086, 09430087, 09430088, 09430089, 09430091, 09430093, 09430094, 09430097, 09430099, 09430100, 09520001, 09520002, 09520003, 09520004, 09520005, 09520006, 09520007, 09520008, 09520009, 09520011, 09520012, 09520013, 09520015, 09520017, 09520018, 09520019, 09520020, 09520021, 09520022, 09520025, 09520026, 09520027, 09520028, 09520029, 09520030, 09520031, 09520032, 09520033, 09520034, 09520035, 09520036, 09520037, 09520038, 09520039, 09520040, 09520041, 09520042, 09520043, 09520044, 09520045, 09520046, 09520047, 09520048, 09520049, 09520050, 10040001, 10040003, 10040004, 10040005, 10040006, 10040007, 10040008, 10040009, 10040010, 10040011, 10040012, 10040013, 10040014, 10040015, 10040016, 10040017, 10040028, 10040029, 10040030, 10040032, 10040033, 10040034, 10040035, 10040036, 10040037, 10040038, 10040039, 10040040, 10040041, 10040042, 10040043, 10040044, 10040045, 10040046, 10040047, 10040048, 10040049, 10040050, 10040051, 10040052, 10040053, 10040054, 10040055, 10040056, 10040057, 10040058, 10040059, 10040062, 10040063, 10040064, 10040065, 10040066, 10040067, 10040068, 10040069, 10040070, 10040071, 10040073, 10040074, 10040075, 10040076, 10040077, 10040078, 10400029, 10400050, 10400056, 10400057, 10400058, 10400059, 10400065, 10370016, 10370017, 10370018, 10370019, 10370020, 10370032, 10370033, 10370034, 10370035, 10370038, 10370052, 10370053, 10370054, 10370056, 10370060, 10370063, 10370078, 10370079, 10370080 , 09380031, 09380029, 09380064, 09380070, 09380089, 09380040, 09380038, 09380085, 09380045, 09380054, 09380060, 09380059, 09380058, 09380048, 09380042, 09380038, 09380069, 09380063, 09380067 & 09380065.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0270-2013 | Class II | Terminated | Voluntary: Firm initiated |