Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2427-2012Class II

12/14 TAPER HIGH OFFSET NECK USE WITH SMITH & NEPHEW STEMS AND 12/14 FEMORAL COMPONENTS ONLY CO-CR, REF...

Smith & Nephew, Inc. / initiated 2012-08-22 / Terminated / root cause: manufacturing process

Smith & Nephew Inc began a recall of 12/14 TAPER HIGH OFFSET NECK USE WITH SMITH & NEPHEW STEMS AND 12/14 FEMORAL COMPONENTS ONLY CO-CR, REF 71352112, Smith & Nephew, Inc. Memphis, TN 38116 USA... on . FDA classified it as Class II and gives the reason as "Modular necks were manufactured out of specification by a vendor; the taper provide is undersized". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2427-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Smith & Nephew, Inc.
Product
12/14 TAPER HIGH OFFSET NECK USE WITH SMITH & NEPHEW STEMS AND 12/14 FEMORAL COMPONENTS ONLY CO-CR, REF 71352112, Smith & Nephew, Inc. Memphis, TN 38116 USA. USE WITH SMITH & NEPHEW STEMS AND 12/14 FEMORAL COMPONENTS.
Product code
MEH Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate
Reason for recall
Modular necks were manufactured out of specification by a vendor; the taper provide is undersized.
Root cause tag
manufacturing process
FDA root cause
Employee error
Action
Smith & Nephew sent an "URGENT:CLASS II RECALL-HIGH OFFSET MODULAR NECK R1213 letter dated August 22, 2012 to all affected customers. The letter identified the product, problem, and the actions to be taken by the customers. Customers were instructed to cease use and distribution of the product and quarantine for return. An Inventory Return Certification Form was included for customers to complete and return via fax to 901-566-7975.
Quantity in commerce
70 units
Distribution
Worldwide Distribution-USA (nationwide) including the states of CA, FL, GA, IL, IN, MA, NC, NJ, NY, OH, OK, PA, TN, TX, and WA, and the country of Canada.
Product codes and lots
Lot Numbers: 12FTM0017A, 12FTM0017B, 12FTM0017C, 12FTM0019A. 12FTM0019B, 12FTM0019C, 12FTM0019D, 12FTM0020A, 12FTM0020B, 12FTM0020C, 12FTM0021A, 12FTM0021B, and 12FTM0021C

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2427-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2427-2012Class IITerminatedVoluntary: Firm initiated