Boule Con-Diff Hematology Control For in vitro diagnostic use as a control to monitor the performance of...
Clinical Diagnostic Solutions began a recall of Boule Con-Diff Hematology Control For in vitro diagnostic use as a control to monitor the performance of multi-parameter hematology instruments on . FDA classified it as Class III and gives the reason as "Clinical Diagnostic Solutions, Inc. in Plantation, FL recalled Boule Con Diff US Tri Pack (Product Part 501-605, Lot #1102-575 and 1102-576; Product Part 501-607, Lot #1102-576; and Product Part 502-012, Lot #1102-574)...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0190-2013
- Classification
- Class III
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Clinical Diagnostic Solutions, Inc.
- Product
- Boule Con-Diff Hematology Control For in vitro diagnostic use as a control to monitor the performance of multi-parameter hematology instruments.
- Product code
- JPK Mixture, Hematology Quality Control
- Reason for recall
- Clinical Diagnostic Solutions, Inc. in Plantation, FL recalled Boule Con Diff US Tri Pack (Product Part 501-605, Lot #1102-575 and 1102-576; Product Part 501-607, Lot #1102-576; and Product Part 502-012, Lot #1102-574) which is used for in vitro diagnostic use as a control to monitor the performance of multi-parameter hematology instruments. The recall was initiated due to the receipt of a complai
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Clinical Diagnostic Solutions, Inc. sent a letter dated March 24, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The firm istructed the accounts to replace the assay sheet they were currently using with the new assay sheets enclosed. For an unused product they would need to discard. To correct the ranges in their instrument, they would need to scan in the new bar codes specific for their instrument type. For any questions customers were instructed to contact CDE Technical Support atl 800-453-3328. For questions regarding this recall call 954-791-1773.
- Quantity in commerce
- 901 vials
- Distribution
- Worldwide Distribution - USA including AL, AR, AZ, CA, CO, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MS, ND, ND, NE, NJ, NY, OH, OK, PA, SC, TN, TX, UT, VA, WI, WV,and WY. Foreign: Brazil and Colombia
- Product codes and lots
- Product Part 501-607, Lot #1102-576
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0189-2013 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-0190-2013 | Class III | Terminated | Voluntary: Firm initiated | ||
| Z-0191-2013 | Class III | Terminated | Voluntary: Firm initiated |