Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2448-2012Class II

Irrigation Catheters (Models IRR-XXX-XX): Model Numbers & Description: IRR-023-10F,CATHETER IRRIGATION 10F...

Applied Medical Resources Corp / initiated 2012-08-09 / Terminated / root cause: manufacturing process

Applied Medical Resources Corp began a recall of Irrigation Catheters (Models IRR-XXX-XX): Model Numbers & Description: IRR-023-10F,CATHETER IRRIGATION 10F 23CM; IRR-080-4F, CATHETER, IRRIGATION 4F 80CM... on . FDA classified it as Class II and gives the reason as "Applied Medical is recalling specific lot numbers of its Vascular Catheters because they may have a potential for packaging particulate matter to reside on the product". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-2448-2012
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Applied Medical Resources Corp
Product
Irrigation Catheters (Models IRR-XXX-XX): Model Numbers & Description: IRR-023-10F,CATHETER IRRIGATION 10F 23CM; IRR-080-4F, CATHETER, IRRIGATION 4F 80CM; IRR-080-6F, CATHETER, IRRIGATION 6F 80CM. Irrigation Catheters are indicated for irrigation in the peripheral vascular system and biliary tree.
Product code
DQY Catheter, Percutaneous
Reason for recall
Applied Medical is recalling specific lot numbers of its Vascular Catheters because they may have a potential for packaging particulate matter to reside on the product.
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
Applied Medical sent an "URGENT: MEDICAL DEVICE RECALL" letter dated August 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Customers were instructed to complete the attached Recall Notification Confirmation Form and to return it to the firm via e-mail to [email protected] or fax to (949) 713-8832. For product return questions, customers were instructed to contact Sales Operations Administrator at (949) 713-8652.For questions regarding this recall contact the firm at (949) 713-8041.
Quantity in commerce
40,238 units for all products in Recall Event
Distribution
Worldwide Distribution-USA (nationwide) and the countries of United Arab Emirates, Austria, Australia, Belgium, Canada, Germany, Ecuador, Estonia, Spain, France, United Kingdom, Hong Kong, Ireland, Israel, Italy, Republic of Korea, Libyan Arab Jamahiriya, Malaysia, Netherlands, New Zealand, Pakistan, Poland, Portugal, Qatar, Serbia, Saudi Arabia, Taiwan, and South Africa.
Product codes and lots
Model Numbers: Lot Numbers IRR-023-10F: 1158230. IRR-080-4F: 1156082, 1170185, 1164648. IRR-080-6F: 1158960.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2448-2012
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2446-2012Class IITerminatedVoluntary: Firm initiated
Z-2447-2012Class IITerminatedVoluntary: Firm initiated
Z-2448-2012Class IITerminatedVoluntary: Firm initiated