Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1836-2013Class II

Eon Mini - 3788 (IPG), Eon Mini system is indicated as an aid in the management of chronic intractable pain...

Abbott Medical / initiated 2012-07-26 / Terminated / root cause: design

St. Jude Medical began a recall of Eon Mini - 3788 (IPG), Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or... on . FDA classified it as Class II and gives the reason as "As part of St. Jude Medical's routine tracking and product monitoring, they have received a total of 214 reports, out of 34,617 Eon Mini IPGs, that lost the ability to communicate or recharge due to an inner battery...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1836-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Eon Mini - 3788 (IPG), Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back, surgery syndrome, intractable low back and leg pain.
Product code
LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
Reason for recall
As part of St. Jude Medical's routine tracking and product monitoring, they have received a total of 214 reports, out of 34,617 Eon Mini IPGs, that lost the ability to communicate or recharge due to an inner battery weld issue resulting in loss of pain relief and subsequent explant. Explant surgery, as with any surgery, presents a risk to patient health. Adverse events associated with an unplanne
Root cause tag
design
FDA root cause
Device Design
Action
St. Jude Medical sent a Important Medical Device Recall Advisory dated July 26, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. To further assist in your patient care we are providing you with a list of all serial numbers we show have been distributed to you. ( See Attachment A ). Further questions please call (972) 309-2154.
Quantity in commerce
34,617
Distribution
Worldwide Distribution--USA (nationwide) including Puerto Rico and the countries of Argentina, Australia, Austria, Belgium, Brazil, canada, Colombia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, India, Ireland, Israel, Italy, Japan, Mexico, Netherlands, Saudi, South Africa, Spain, Sweden, Switzerland, Turkey and United Kingdom.
Product codes and lots
Multiple lots recalled are the ones with supplier's inner battery lots of 326-508

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1836-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1836-2013Class IITerminatedVoluntary: Firm initiated