Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1738-2013Class II

Eon - 3701 (Wall Charger) and 3711 (Portable Charger)

Abbott Medical / initiated 2012-07-26 / Terminated / root cause: design

St. Jude Medical began a recall of Eon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of... on . FDA classified it as Class II and gives the reason as "St. Jude Medical has informed your doctor that a number of cases have been reported in which discomfort associated with heating occurred at the device site while patients were using the Charging System to charge their...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1738-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Eon - 3701 (Wall Charger) and 3711 (Portable Charger). Product Usage: Eon: Eon system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back surgery syndrome, intractable low back and leg pain. Eon Mini: Eon Mini system is indicated as an aid in the management of chronic intractable pain of the trunk and/or limbs including unilateral or bilateral pain associated with the following: failed back surgery syndrome, intractable low back and leg pain.
Product code
LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
Reason for recall
St. Jude Medical has informed your doctor that a number of cases have been reported in which discomfort associated with heating occurred at the device site while patients were using the Charging System to charge their spinal cord stimulator. In three cases, patients received a burn to the skin (one second-degree and two first-degree burns) at the implant site. As of June 30, 2012, St. Jude Medica
Root cause tag
design
FDA root cause
Device Design
Action
ST. Jude Medical sent an Important Medical Device Safety Information dated July 26, 2012 to affected customers. The letter identifed the affected product, problem and actions to be taken. For questions or concerns contact your ST. Jude Medical representative or St. Jude Medical Neuromodulation Division at 1-866-240-6741.
Quantity in commerce
70,638 (combined total for Eon and Eon mini systems)
Distribution
Worldwide Distribution - USA Nationwide Including Puerto Rico and countries of: Argentina, Australia, Belgium, Brazil, Canada, China, Chile, Colombia, Ireland, Israel, India, Malaysia, Mexico, Saudi Arabia, Singapore, South Africa, South Korea, Switzerland and United Kingdom.
Product codes and lots
All serial numbers manufactured

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1738-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1738-2013Class IITerminatedVoluntary: Firm initiated
Z-1739-2013Class IITerminatedVoluntary: Firm initiated