Z-0110-2013Class III

Synchron Systems Synchron Calibrator Multi Calibrator Diskettes, Part Number 442600. The SYNCHRON...

Siemens Healthcare Diagnostics Inc. / initiated 2011-12-15 / Terminated / root cause: labeling and UDI

Beckman Coulter Inc. began a recall of Synchron Systems Synchron Calibrator Multi Calibrator Diskettes, Part Number 442600. The SYNCHRON MULTI-CALIBRATOR, used in conjunction with SYNCHRON reagents... on . FDA classified it as Class III and gives the reason as "The recall was initiated because Beckman Coulter has confirmed a mislabeling error, some Synchron Systems Multi Calibrator Diskettes (REF 442600, lot M002642G) may actually be Thyroxine (T4) Reagent Calibrator Diskettes". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0110-2013
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Synchron Systems Synchron Calibrator Multi Calibrator Diskettes, Part Number 442600. The SYNCHRON MULTI-CALIBRATOR, used in conjunction with SYNCHRON reagents, is intended for use on SYNCHRON Systems for the calibration of Albumin, Blood Urea Nitrogen (Urea), Calcium, Cholesterol, Glucose, Inorganic Phosphorus, Lactate,Magnesium, Total Protein, Triglycerides, and Uric Acid.
Product code
JIX Calibrator, Multi-Analyte Mixture
Reason for recall
The recall was initiated because Beckman Coulter has confirmed a mislabeling error, some Synchron Systems Multi Calibrator Diskettes (REF 442600, lot M002642G) may actually be Thyroxine (T4) Reagent Calibrator Diskettes.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
Beckman Coulter sent an Urgent Product Correction letter dated December 15, 2011, withan attached Fax Back Response Form to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. If customers are currently using Multi Calibrator lot M002642, then they were instructed to verified diskette and discard all Multi Calibrator diskettes lot M002642 and M002642G without a dark green dot. Customers were instructed to complete and return the enclosed Response Form within 10 days. Customers were instructed to contact Beckman Coulter Support Center at http:www.beckmancoulter.com/customersupport/support/ or call (800) 854-3633 in the US and Canada. Outside of the US or Canada, customers were instructed to contact their local Beckman Coulter Representatives. For questions regarding this recall call 714-961-4483.
Quantity in commerce
5746 units total (2729 units in US)
Distribution
Worldwide distribution: USA (Nationwide) and Internationally to: Afghanistan Angola Australia Bangladesh Belgium Bhutan Brazil Brunei Darussalam Bulgaria Canada China Colombia Costa Rica Cote d'Ivoire Czech Republic Denmark Ecuador Egypt El Salvador Eritrea Ethiopia Finland France French Polynesia Germany Ghana Greece Guatemala Hong Kong India Indonesia Ireland Israel Italy Japan Jordan Kenya Kuwait Lebanon Lesotho Libyan Arab Jamahiriya Lithuania Macao Malawi Malaysia Mexico Morocco Namibia Netherlands New Zealand Nigeria Oman Pakistan Panama Paraguay Philippines Poland Puerto Rico Qatar Reunion Romania Russian Federation Rwanda Saudi Arabia Serbia Singapore Slovakia South Africa Spain Swaziland Sweden Switzerland Taiwan Thailand Tunisia Turkey Uganda United Arab Emirates United Kingdom Viet Nam Zambia Zimbabwe
Product codes and lots
Lot M002642

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0110-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0110-2013Class IIITerminatedVoluntary: Firm initiated